What does the Midmark M11 Error Code E011 Mean?
Error E011: Hardware malfunction or corrupted software.
Probable causes of Midmark M11 Error Code E011
PC Control Board has had a software/hardware failure.
Solution: Unplug the unit and wait 60 seconds, plug back in and press On/Standby button. If the display and buttons have not lit up, then it is probably the PC Board that needs replaced. This is not an end user repair.
If you have any problems or are unsure, please use our FREE TECH SUPPORT to contact us.
View our Midmark M11 “Old Style” Error Code E011 Troubleshooting Video Below for More Help!