Statim G4 5000 Check Valve Replacement / Service
Autoclave maintenance is important. Here at SAS, we state the importance of maintenance as much as possible. And we make it our mission to provide various services to ensure your autoclave's longevity.
So in this video, we'll go over maintenance for the Statim G4 5000 steam sterilizer. Specifically, the replacement of the Check Valve. The Check Valve is one of the parts in need of replacement and/or maintenance after 2 years.
You can find the Statim G4 5000 Check Valve Repair Kit on our website using the link here:
https://statimusa.com/product/scican-statim-g4-5000-check-valve-oem-01-101627s/
If you have any questions, feel free to call us at 704-966-1650, or you can check out our different programs using the links below.
And if you have any questions about this issue or anything else, you can call us at 704-966-1650, option 3 for free tech support, or check us out using the links below:
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#sterilization #autoclave #scican
Statim G4 2000 Cassette Maintenance / Service
Autoclave maintenance is important. Here at SAS, we state the importance of maintenance as much as possible. And we make it our mission to provide various services to ensure your autoclave's longevity.
So in this video, we'll go over one of the maintenance steps for the Statim G4 2000 steam sterilizer. Specifically, the maintenance and cleaning of the cassette. We have Ian here to walk you through how to disassemble the cassette and what tools to use for proper cleaning.
If you'd like to purchase a Statim G4 2000 Cassette Seal and Lube kit, you can find it on our site here: https://statimusa.com/product/new-scican-statim-g4-2000-cassette-seal-and-lube-kit-oem-01-100028s/
If you'd like to purchase a Statim G4 2000 Cassette Lid, you can find them on our site here: https://statimusa.com/product/scican-statim-g4-2000-cassette-lid-oem-01-112409s/
If you'd like to purchase a Statim G4 2000 Cassette Tray and Rack kit, you can find them on our site here: https://statimusa.com/product/scican-statim-g4-2000-cassette-tray-with-rack-oem-01-11240s/
If you have any questions, feel free to call us at 704-966-1650, or you can check out our different programs using the links below.
And if you have any questions about this issue or anything else, you can call us at 704-966-1650, option 3 for free tech support, or check us out using the links below:
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#maintenance #autoclave #scican
Statim G4 2000 Top Cover Replacement / Service
Autoclave maintenance is important. Here at SAS, we state the importance of maintenance as much as possible. And we make it our mission to provide various services to ensure your autoclave's longevity.
So in this video, we'll go over one of the maintenance steps for the Statim G4 2000 steam sterilizer. Specifically, the removal and subsequent replacement of the top cover.
If you have any questions, feel free to call us at 704-966-1650, or you can check out our different programs using the links below.
And if you have any questions about this issue or anything else, you can call us at 704-966-1650, option 3 for free tech support, or check us out using the links below:
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#maintenance #autoclave #scican
Statim G4 2000 Biological Filter Replacement / Service #shorts
Autoclave maintenance is important. Here at SAS, we state the importance of maintenance as much as possible. And we make it our mission to provide various services to ensure your autoclave's longevity.
So in this short, we'll go over one of the maintenance steps for the Statim G4 2000 steam sterilizer. Specifically, the replacement of the biological filter.
We also have the Statim G4 2000 Biological Filter on our website here: https://statimusa.com/product/scican-statim-g4-2000-5000-biological-filter-oem-01-102119s/
If you have any questions, feel free to call us at 704-966-1650, or you can check out our different programs using the links below.
And if you have any questions about this issue or anything else, you can call us at 704-966-1650, option 3 for free tech support, or check us out using the links below:
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Timestamps:
00:00 Introduction
00:12 Biological FIlter Replacement
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#sterilization #shorts #scican
Statim G4 2000 Compressor Filter Replacement / Service
Autoclave maintenance is important. Here at SAS, we state the importance of maintenance as much as possible. And we make it our mission to provide various services to ensure your autoclave's longevity.
So in this video, we'll go over one of the maintenance steps for the Statim G4 2000 steam sterilizer. Specifically, the replacement of the compressor filter.
We also have the Statim G4 2000 Compressor Filter on our website here: https://statimusa.com/product/scican-statim-2000-compressor-filter-rpi-part-scf002-oem-part-01-100207s/
If you have any questions, feel free to call us at 704-966-1650, or you can check out our different programs using the links below.
And if you have any questions about this issue or anything else, you can call us at 704-966-1650, option 3 for free tech support, or check us out using the links below:
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#maintenance #autoclave #scican
Statim G4 5000 Solenoid Valve Cleaning / Service
Autoclave maintenance is important. Here at SAS, we state the importance of maintenance as much as possible. And we make it our mission to provide various services to ensure your autoclave's longevity.
So in this video, we'll go over maintenance for the Statim G4 5000 steam sterilizer. Specifically, the replacement of the Solenoid Valve. The Solenoid Valve is one of the parts in need of replacement and/or maintenance after 2 years.
You can find the Statim G4 5000 Solenoid Valve Complete Repair Kit on our website using the link here:
https://statimusa.com/product/scican-statim-g4-5000-solenoid-valve-complete-oem-01-101628s/
If you have any questions, feel free to call us at 704-966-1650, or you can check out our different programs using the links below.
And if you have any questions about this issue or anything else, you can call us at 704-966-1650, option 3 for free tech support, or check us out using the links below:
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#sterilization #autoclave #scican
Statim G4 2000 Solenoid Valve Cleaning / Service
Autoclave maintenance is important. Here at SAS, we state the importance of maintenance as much as possible. And we make it our mission to provide various services to ensure your autoclave's longevity.
So in this video, we'll go over maintenance for the Statim G4 2000 steam sterilizer. Specifically, the replacement of the Solenoid Valve. The Solenoid Valve is one of the parts in need of replacement and/or maintenance after 2 years.
You can find the Statim G4 2000 Solenoid Valve Complete Repair Kit on our website using the link here:
https://statimusa.com/product/scican-statim-g4-2000-solenoid-valve-complete-oem-01-100557s/
If you have any questions, feel free to call us at 704-966-1650, or you can check out our different programs using the links below.
And if you have any questions about this issue or anything else, you can call us at 704-966-1650, option 3 for free tech support, or check us out using the links below:
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#sterilization #autoclave #scican
StatClave G4 Cycle Fault 77 Overview
In this video, we'll go over Cycle Fault 77 on the StatClave G4 autoclave. Cycle Fault 77 means the there's an issue with the vacuum system in the unit. So, we brought in our technician, Jacob, to show how to go about diagnosing and solving this issue.
If you have any questions about SciCan, their StatClave G4 or anything else, you can contact us at 704-966-1650 option 3 for our Free Tech Support line, or you can find us using the links to our service programs below.
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#scican #sterilization #autoclave
StatClave G4 Cycle Fault 73 Overview
In this video, we'll go over Cycle Fault 73 on the StatClave G4 autoclave. Cycle Fault 73 means the there's an issue with the vacuum system in the unit. So, we brought in our technician, Jacob, to show how to go about diagnosing and solving this issue.
If you have any questions about SciCan, their StatClave G4 or anything else, you can contact us at 704-966-1650 option 3 for our Free Tech Support line, or you can find us using the links to our service programs below.
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#scican #sterilization #autoclave
StatClave G4 Cycle Fault 43 Overview
In this video, we'll go over Cycle Fault 43 on the StatClave G4 autoclave. Cycle Fault 43 means the there's an issue with the automatic filling machine. So, we brought in our technician, Jacob, to show how to go about diagnosing and solving this issue.
If you have any questions about SciCan, their StatClave G4 or anything else, you can contact us at 704-966-1650 option 3 for our Free Tech Support line, or you can find us using the links to our service programs below.
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#scican #sterilization #autoclave
SciCan Statim 5000 Cycle Fault 3
Troubleshooting errors on your Statim 5000 autoclave can seem like a daunting task, luckily, SAS is here to do what we do best. So in this video, we'll go over Cycle Fault 3 on the SciCan Statim 5000. This Cycle Fault 3 consists of multiple troubleshooting options, only one of which is able to be done by the end-user. Every other repair is meant to be serviced by a service technician.
So in this video, we'll walk you through how the end-user can go about their repair, what the service technician can do to help, and where to go if the problem persists beyond the end-user repair.
If you have any questions about SciCan, their StatClave G4 or anything else, you can contact us at 704-966-1650 option 3 for our Free Tech Support line, or you can find us using the links to our service programs below.
The Statim Double Ended Spanner Wrench is available here on our site:
https://statimusa.com/product/double-ended-spanner-wrench-rpi-part-rpt501/
Our SciCan Playlist is available here on our YouTube Channel:
https://www.youtube.com/playlist?list=PLEvgTwUHjK9v7ZPIp7v-8slJoVagW9EfX
Our video on replacing the Statim 5000 Cassette Seal is available here:
https://youtu.be/hNXYjoCWBko
Our video on replacing the Statim 5000 Top Cover is available here:
https://youtu.be/sXA-1xmBexg
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#scican #sterilization #autoclave
StatClave G4 Cycle Fault 44 Overview
In this video, one of our technicians, Jacob, will go over the Cycle Fault 44 error on the StatClave G4. Usually, the Cycle Fault 44 happens when the water float inside of the Venturi reservoir is activated constantly. Usually caused by an overfull venturi reservoir.
So in this video, Jacob will walk you through how to go about fixing this error should it ever occur for an end-user.
If you have any questions about SciCan, their StatClave G4 or anything else, you can contact us at 704-966-1650 option 3 for our Free Tech Support line, or you can find us using the links to our service programs below.
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#scican #sterilization #autoclave
StatClave G4 Cycle Fault 16 Overview
In this video, one of our technicians, Jacob, will go over the Cycle Fault 16 error on the StatClave G4. Usually, the Cycle Fault 16 has to do with the the inner parts of the unit getting too hot, or issues with the floater and water level inside of the reservoir.
So in this video, Jacob will walk you through how to go about fixing this error should it ever occur for an end-user.
If you have any questions about SciCan, their StatClave G4 or anything else, you can contact us at 704-966-1650 option 3 for our Free Tech Support line, or you can find us using the links to our service programs below.
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#scican #sterilization #autoclave
StatClave G4 Cycle Fault 10 Overview
In this video, one of our technicians, Jacob, will go over the Cycle Fault 10 error on the StatClave G4. Usually, the Cycle Fault 10 has to do with the exhaust lines on your unit being kinked, pinched or improperly flowing. As the exhaust tubes need to sit properly to go down with the flow of gravity.
So Jacob will go over Cycle fault 10, what to do when you get it, and where to go if the problem persists.
If you have any questions about SciCan, their StatClave G4 or anything else, you can contact us at 704-966-1650 option 3 for our Free Tech Support line, or you can find us using the links to our service programs below.
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#scican #sterilization #autoclave
StatClave G4 Cycle Fault 4 Overview
In this video, one of our technicians, Jacob, will go over the Cycle Fault 4 error on the StatClave G4. Usually, the Cycle Fault 4 means that there could be a problem with the door gasket.
So Jacob will go over Cycle fault 4, what to do when you get it, and where to go if the problem persists.
If you have any questions about SciCan, their StatClave G4 or anything else, you can contact us at 704-966-1650 option 3 for our Free Tech Support line, or you can find us using the links to our service programs below.
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#scican #sterilization #autoclave
SciCan Bravo G4 Error Code E901
In this video, we're covering one of the many error codes for the SciCan Bravo G4. This time, we're going over Error Code E901. Error Code E901 relates to a failure during the vacuum test cycle of your Bravo G4 during the standby phase. So in this video, Andrew will go over what the error is and what you can do if it ever appears.
You can find the Bravo G4 Door Gasket here on our website:
https://statimusa.com/product/scican-bravo-17-21-door-gasket-oem-97400145/
If you have any questions about this or anything else, you can contact us at 704-966-1650, or you can find us using the links to our service programs below.
You can find the SciCan Bravo G4 unit using the link below:
https://statimusa.com/?orderby=price-desc&paged=1&s=bravo+g4&product_cat=0&post_type=product
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Timestamps:
00:00 Intro
00:09 Code Description
00:23 What to Do
01:24 Outro
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#scican #sterilization #autoclave
SciCan Bravo G4 Error Code E900
In this video, we're covering one of the many error codes for the SciCan Bravo G4. This time, we're going over Error Code E900. Error Code E900 relates to a failure during the test cycle of your Bravo G4. So in this video, Andrew will go over what the error is and what you can do if it ever appears.
You can find the Bravo G4 Door Gasket here on our website:
https://statimusa.com/product/scican-bravo-17-21-door-gasket-oem-97400145/
If you have any questions about this or anything else, you can contact us at 704-966-1650, or you can find us using the links to our service programs below.
You can find the SciCan Bravo G4 unit using the link below:
https://statimusa.com/?orderby=price-desc&paged=1&s=bravo+g4&product_cat=0&post_type=product
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Timestamps:
00:00 Intro
00:10 Code Description
00:23 What to Do
01:11 Outro
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#scican #sterilization #autoclave
SciCan Bravo G4 Error Code E141
In this video, we're covering one of the many error codes for the SciCan Bravo G4. This time, we're going over Error Code E141. Error Code E141 is an issue with the unit having the incorrect firmware version, and is in need of a fix to set it right. With the necessary resolution being to contact a qualified technician for help, because any repairs are NOT meant for end-users.
If you have any questions about this or anything else, you can contact us at 704-966-1650, or you can find us using the links to our service programs below.
You can find the SciCan Bravo G4 unit using the link below:
https://statimusa.com/?orderby=price-desc&paged=1&s=bravo+g4&product_cat=0&post_type=product
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Timestamps:
00:00 Intro
00:10 Code Description
00:26 What to Do
00:38 Outro
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#scican #sterilization #autoclave
SciCan Bravo G4 Error Code E126
In this video, we're covering one of the many error codes for the SciCan Bravo G4. This time, we're going over Error Code E126. Error Code E126 is an error that relates to an update in the cloud firmware of your unit. So we brought in Andrew to explain the error, what you can do and where to go if the error persists.
If you have any questions about this or anything else, you can contact us at 704-966-1650, or you can find us using the links to our service programs below.
You can find the SciCan Bravo G4 unit using the link below:
https://statimusa.com/?orderby=price-desc&paged=1&s=bravo+g4&product_cat=0&post_type=product
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Timestamps:
00:00 Intro
00:08 Code Description
00:19 What to Do
00:39 Outro
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#scican #sterilization #autoclave
SciCan Bravo G4 Error Code E061
In this video, we're covering one of the many error codes for the SciCan Bravo G4. This time, we're going over Error Code E061. Error Code E061 is an issue with the connection of your unit to the internet through WiFi connection. So our technician, Andrew, is here to tell you where to go from the error code to solve the problem.
If you have any questions about this or anything else, you can contact us at 704-966-1650, or you can find us using the links to our service programs below.
You can find the SciCan Bravo G4 unit using the link below:
https://statimusa.com/?orderby=price-desc&paged=1&s=bravo+g4&product_cat=0&post_type=product
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Timestamps:
00:00 Intro
00:08 Code Description
00:19 What to Do
00:39 Outro
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#scican #sterilization #autoclave
Statim G4 5000 Check Valve Replacement / Service
Statim G4 5000 Check Valve Replacement / Service
Autoclave maintenance is important. Here at SAS, we state the importance of maintenance as much as possible. And we make it our mission to provide various services to ensure your autoclave's longevity.
So in this video, we'll go over maintenance for the Statim G4 5000 steam sterilizer. Specifically, the replacement of the Check Valve. The Check Valve is one of the parts in need of replacement and/or maintenance after 2 years.
You can find the Statim G4 5000 Check Valve Repair Kit on our website using the link here:
https://statimusa.com/product/scican-statim-g4-5000-check-valve-oem-01-101627s/
If you have any questions, feel free to call us at 704-966-1650, or you can check out our different programs using the links below.
And if you have any questions about this issue or anything else, you can call us at 704-966-1650, option 3 for free tech support, or check us out using the links below:
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#sterilization #autoclave #scican
Statim G4 2000 Cassette Maintenance / Service
Autoclave maintenance is important. Here at SAS, we state the importance of maintenance as much as possible. And we make it our mission to provide various services to ensure your autoclave's longevity.
So in this video, we'll go over one of the maintenance steps for the Statim G4 2000 steam sterilizer. Specifically, the maintenance and cleaning of the cassette. We have Ian here to walk you through how to disassemble the cassette and what tools to use for proper cleaning.
If you'd like to purchase a Statim G4 2000 Cassette Seal and Lube kit, you can find it on our site here: https://statimusa.com/product/new-scican-statim-g4-2000-cassette-seal-and-lube-kit-oem-01-100028s/
If you'd like to purchase a Statim G4 2000 Cassette Lid, you can find them on our site here: https://statimusa.com/product/scican-statim-g4-2000-cassette-lid-oem-01-112409s/
If you'd like to purchase a Statim G4 2000 Cassette Tray and Rack kit, you can find them on our site here: https://statimusa.com/product/scican-statim-g4-2000-cassette-tray-with-rack-oem-01-11240s/
If you have any questions, feel free to call us at 704-966-1650, or you can check out our different programs using the links below.
And if you have any questions about this issue or anything else, you can call us at 704-966-1650, option 3 for free tech support, or check us out using the links below:
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#maintenance #autoclave #scican
Statim G4 2000 Top Cover Replacement / Service
Autoclave maintenance is important. Here at SAS, we state the importance of maintenance as much as possible. And we make it our mission to provide various services to ensure your autoclave's longevity.
So in this video, we'll go over one of the maintenance steps for the Statim G4 2000 steam sterilizer. Specifically, the removal and subsequent replacement of the top cover.
If you have any questions, feel free to call us at 704-966-1650, or you can check out our different programs using the links below.
And if you have any questions about this issue or anything else, you can call us at 704-966-1650, option 3 for free tech support, or check us out using the links below:
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#maintenance #autoclave #scican
Statim G4 2000 Biological Filter Replacement / Service #shorts
Autoclave maintenance is important. Here at SAS, we state the importance of maintenance as much as possible. And we make it our mission to provide various services to ensure your autoclave's longevity.
So in this short, we'll go over one of the maintenance steps for the Statim G4 2000 steam sterilizer. Specifically, the replacement of the biological filter.
We also have the Statim G4 2000 Biological Filter on our website here: https://statimusa.com/product/scican-statim-g4-2000-5000-biological-filter-oem-01-102119s/
If you have any questions, feel free to call us at 704-966-1650, or you can check out our different programs using the links below.
And if you have any questions about this issue or anything else, you can call us at 704-966-1650, option 3 for free tech support, or check us out using the links below:
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Timestamps:
00:00 Introduction
00:12 Biological FIlter Replacement
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#sterilization #shorts #scican
Statim G4 2000 Compressor Filter Replacement / Service
Autoclave maintenance is important. Here at SAS, we state the importance of maintenance as much as possible. And we make it our mission to provide various services to ensure your autoclave's longevity.
So in this video, we'll go over one of the maintenance steps for the Statim G4 2000 steam sterilizer. Specifically, the replacement of the compressor filter.
We also have the Statim G4 2000 Compressor Filter on our website here: https://statimusa.com/product/scican-statim-2000-compressor-filter-rpi-part-scf002-oem-part-01-100207s/
If you have any questions, feel free to call us at 704-966-1650, or you can check out our different programs using the links below.
And if you have any questions about this issue or anything else, you can call us at 704-966-1650, option 3 for free tech support, or check us out using the links below:
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#maintenance #autoclave #scican
Statim G4 5000 Solenoid Valve Cleaning / Service
Autoclave maintenance is important. Here at SAS, we state the importance of maintenance as much as possible. And we make it our mission to provide various services to ensure your autoclave's longevity.
So in this video, we'll go over maintenance for the Statim G4 5000 steam sterilizer. Specifically, the replacement of the Solenoid Valve. The Solenoid Valve is one of the parts in need of replacement and/or maintenance after 2 years.
You can find the Statim G4 5000 Solenoid Valve Complete Repair Kit on our website using the link here:
https://statimusa.com/product/scican-statim-g4-5000-solenoid-valve-complete-oem-01-101628s/
If you have any questions, feel free to call us at 704-966-1650, or you can check out our different programs using the links below.
And if you have any questions about this issue or anything else, you can call us at 704-966-1650, option 3 for free tech support, or check us out using the links below:
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#sterilization #autoclave #scican
Statim G4 2000 Solenoid Valve Cleaning / Service
Autoclave maintenance is important. Here at SAS, we state the importance of maintenance as much as possible. And we make it our mission to provide various services to ensure your autoclave's longevity.
So in this video, we'll go over maintenance for the Statim G4 2000 steam sterilizer. Specifically, the replacement of the Solenoid Valve. The Solenoid Valve is one of the parts in need of replacement and/or maintenance after 2 years.
You can find the Statim G4 2000 Solenoid Valve Complete Repair Kit on our website using the link here:
https://statimusa.com/product/scican-statim-g4-2000-solenoid-valve-complete-oem-01-100557s/
If you have any questions, feel free to call us at 704-966-1650, or you can check out our different programs using the links below.
And if you have any questions about this issue or anything else, you can call us at 704-966-1650, option 3 for free tech support, or check us out using the links below:
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#sterilization #autoclave #scican
StatClave G4 Cycle Fault 77 Overview
In this video, we'll go over Cycle Fault 77 on the StatClave G4 autoclave. Cycle Fault 77 means the there's an issue with the vacuum system in the unit. So, we brought in our technician, Jacob, to show how to go about diagnosing and solving this issue.
If you have any questions about SciCan, their StatClave G4 or anything else, you can contact us at 704-966-1650 option 3 for our Free Tech Support line, or you can find us using the links to our service programs below.
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#scican #sterilization #autoclave
StatClave G4 Cycle Fault 73 Overview
In this video, we'll go over Cycle Fault 73 on the StatClave G4 autoclave. Cycle Fault 73 means the there's an issue with the vacuum system in the unit. So, we brought in our technician, Jacob, to show how to go about diagnosing and solving this issue.
If you have any questions about SciCan, their StatClave G4 or anything else, you can contact us at 704-966-1650 option 3 for our Free Tech Support line, or you can find us using the links to our service programs below.
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#scican #sterilization #autoclave
StatClave G4 Cycle Fault 43 Overview
In this video, we'll go over Cycle Fault 43 on the StatClave G4 autoclave. Cycle Fault 43 means the there's an issue with the automatic filling machine. So, we brought in our technician, Jacob, to show how to go about diagnosing and solving this issue.
If you have any questions about SciCan, their StatClave G4 or anything else, you can contact us at 704-966-1650 option 3 for our Free Tech Support line, or you can find us using the links to our service programs below.
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#scican #sterilization #autoclave
SciCan Statim 5000 Cycle Fault 3
Troubleshooting errors on your Statim 5000 autoclave can seem like a daunting task, luckily, SAS is here to do what we do best. So in this video, we'll go over Cycle Fault 3 on the SciCan Statim 5000. This Cycle Fault 3 consists of multiple troubleshooting options, only one of which is able to be done by the end-user. Every other repair is meant to be serviced by a service technician.
So in this video, we'll walk you through how the end-user can go about their repair, what the service technician can do to help, and where to go if the problem persists beyond the end-user repair.
If you have any questions about SciCan, their StatClave G4 or anything else, you can contact us at 704-966-1650 option 3 for our Free Tech Support line, or you can find us using the links to our service programs below.
The Statim Double Ended Spanner Wrench is available here on our site:
https://statimusa.com/product/double-ended-spanner-wrench-rpi-part-rpt501/
Our SciCan Playlist is available here on our YouTube Channel:
https://www.youtube.com/playlist?list=PLEvgTwUHjK9v7ZPIp7v-8slJoVagW9EfX
Our video on replacing the Statim 5000 Cassette Seal is available here:
https://youtu.be/hNXYjoCWBko
Our video on replacing the Statim 5000 Top Cover is available here:
https://youtu.be/sXA-1xmBexg
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#scican #sterilization #autoclave
StatClave G4 Cycle Fault 44 Overview
In this video, one of our technicians, Jacob, will go over the Cycle Fault 44 error on the StatClave G4. Usually, the Cycle Fault 44 happens when the water float inside of the Venturi reservoir is activated constantly. Usually caused by an overfull venturi reservoir.
So in this video, Jacob will walk you through how to go about fixing this error should it ever occur for an end-user.
If you have any questions about SciCan, their StatClave G4 or anything else, you can contact us at 704-966-1650 option 3 for our Free Tech Support line, or you can find us using the links to our service programs below.
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#scican #sterilization #autoclave
StatClave G4 Cycle Fault 16 Overview
In this video, one of our technicians, Jacob, will go over the Cycle Fault 16 error on the StatClave G4. Usually, the Cycle Fault 16 has to do with the the inner parts of the unit getting too hot, or issues with the floater and water level inside of the reservoir.
So in this video, Jacob will walk you through how to go about fixing this error should it ever occur for an end-user.
If you have any questions about SciCan, their StatClave G4 or anything else, you can contact us at 704-966-1650 option 3 for our Free Tech Support line, or you can find us using the links to our service programs below.
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#scican #sterilization #autoclave
StatClave G4 Cycle Fault 10 Overview
In this video, one of our technicians, Jacob, will go over the Cycle Fault 10 error on the StatClave G4. Usually, the Cycle Fault 10 has to do with the exhaust lines on your unit being kinked, pinched or improperly flowing. As the exhaust tubes need to sit properly to go down with the flow of gravity.
So Jacob will go over Cycle fault 10, what to do when you get it, and where to go if the problem persists.
If you have any questions about SciCan, their StatClave G4 or anything else, you can contact us at 704-966-1650 option 3 for our Free Tech Support line, or you can find us using the links to our service programs below.
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#scican #sterilization #autoclave
StatClave G4 Cycle Fault 4 Overview
In this video, one of our technicians, Jacob, will go over the Cycle Fault 4 error on the StatClave G4. Usually, the Cycle Fault 4 means that there could be a problem with the door gasket.
So Jacob will go over Cycle fault 4, what to do when you get it, and where to go if the problem persists.
If you have any questions about SciCan, their StatClave G4 or anything else, you can contact us at 704-966-1650 option 3 for our Free Tech Support line, or you can find us using the links to our service programs below.
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#scican #sterilization #autoclave
SciCan Bravo G4 Error Code E901
In this video, we're covering one of the many error codes for the SciCan Bravo G4. This time, we're going over Error Code E901. Error Code E901 relates to a failure during the vacuum test cycle of your Bravo G4 during the standby phase. So in this video, Andrew will go over what the error is and what you can do if it ever appears.
You can find the Bravo G4 Door Gasket here on our website:
https://statimusa.com/product/scican-bravo-17-21-door-gasket-oem-97400145/
If you have any questions about this or anything else, you can contact us at 704-966-1650, or you can find us using the links to our service programs below.
You can find the SciCan Bravo G4 unit using the link below:
https://statimusa.com/?orderby=price-desc&paged=1&s=bravo+g4&product_cat=0&post_type=product
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Timestamps:
00:00 Intro
00:09 Code Description
00:23 What to Do
01:24 Outro
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#scican #sterilization #autoclave
SciCan Bravo G4 Error Code E900
In this video, we're covering one of the many error codes for the SciCan Bravo G4. This time, we're going over Error Code E900. Error Code E900 relates to a failure during the test cycle of your Bravo G4. So in this video, Andrew will go over what the error is and what you can do if it ever appears.
You can find the Bravo G4 Door Gasket here on our website:
https://statimusa.com/product/scican-bravo-17-21-door-gasket-oem-97400145/
If you have any questions about this or anything else, you can contact us at 704-966-1650, or you can find us using the links to our service programs below.
You can find the SciCan Bravo G4 unit using the link below:
https://statimusa.com/?orderby=price-desc&paged=1&s=bravo+g4&product_cat=0&post_type=product
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Timestamps:
00:00 Intro
00:10 Code Description
00:23 What to Do
01:11 Outro
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#scican #sterilization #autoclave
SciCan Bravo G4 Error Code E141
In this video, we're covering one of the many error codes for the SciCan Bravo G4. This time, we're going over Error Code E141. Error Code E141 is an issue with the unit having the incorrect firmware version, and is in need of a fix to set it right. With the necessary resolution being to contact a qualified technician for help, because any repairs are NOT meant for end-users.
If you have any questions about this or anything else, you can contact us at 704-966-1650, or you can find us using the links to our service programs below.
You can find the SciCan Bravo G4 unit using the link below:
https://statimusa.com/?orderby=price-desc&paged=1&s=bravo+g4&product_cat=0&post_type=product
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Timestamps:
00:00 Intro
00:10 Code Description
00:26 What to Do
00:38 Outro
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#scican #sterilization #autoclave
SciCan Bravo G4 Error Code E126
In this video, we're covering one of the many error codes for the SciCan Bravo G4. This time, we're going over Error Code E126. Error Code E126 is an error that relates to an update in the cloud firmware of your unit. So we brought in Andrew to explain the error, what you can do and where to go if the error persists.
If you have any questions about this or anything else, you can contact us at 704-966-1650, or you can find us using the links to our service programs below.
You can find the SciCan Bravo G4 unit using the link below:
https://statimusa.com/?orderby=price-desc&paged=1&s=bravo+g4&product_cat=0&post_type=product
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Timestamps:
00:00 Intro
00:08 Code Description
00:19 What to Do
00:39 Outro
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#scican #sterilization #autoclave
SciCan Bravo G4 Error Code E061
In this video, we're covering one of the many error codes for the SciCan Bravo G4. This time, we're going over Error Code E061. Error Code E061 is an issue with the connection of your unit to the internet through WiFi connection. So our technician, Andrew, is here to tell you where to go from the error code to solve the problem.
If you have any questions about this or anything else, you can contact us at 704-966-1650, or you can find us using the links to our service programs below.
You can find the SciCan Bravo G4 unit using the link below:
https://statimusa.com/?orderby=price-desc&paged=1&s=bravo+g4&product_cat=0&post_type=product
Our Free Tech Support Program
https://statimusa.com/free-autoclave-technical-support/?v=2e5df5aa3470
Our Free Virtual Technician Support
https://statimusa.com/stanley-sterilizer-your-virtual-autoclave-technician/?v=2e5df5aa3470
Our Free Preventative Maintenance Program
https://statimusa.com/connect-with-us/?v=2e5df5aa3470
Timestamps:
00:00 Intro
00:08 Code Description
00:19 What to Do
00:39 Outro
Free Use Music by OlexandrMusic
https://www.olexandrignatov.com/freemusic
https://soundcloud.com/olexandrmusic
https://www.youtube.com/olexandrmusic
https://www.olexandrignatov.com
#scican #sterilization #autoclave
Statim G4 5000 Check Valve Replacement / Service