This article explains the possible causes and solutions to a Midmark M9-M11 New Style autoclave Error Code C102.

Midmark M9/M11 Error Code C102

In this article, we’re going to explain one of the many error codes that can affect your Midmark M9/Midmark M11 autoclave. This article will focus specifically on error code C102. We’ll tell you about the problem, what displays when the error occurs, its possible causes, and the course of action to take to resolve said error.

 

However if you’re more of a visual learner we have a video on this error code available here. 

And if you just want to talk to someone, call our Free Tech Support at 704-966-1650 Option 3.

Display/Symptom

C102: FILL MODE STOP PRESSED

Causes, Checks and Actions

Possible Cause 1: STOP button was pressed.
  • Check: No Check Necessary.
  • Action: Press STOP button and initiate a new cycle.

Possible Cause 2: Touch pad malfunctioning. 
  • Check: Perform Key test in service diagnostic.
  • Action: Replace touch pad.

A replacement Touch Pad is available to buy here. 