Using the Machine

These control operations (reported below) are performed to check if the machine works properly and should be performed when the machine installation is completed. 

1) Open the tap that supplies water to the machine. 


WARNING! The water should NOT flow into the chamber; otherwise the water loading solenoid valves are dirty or blocked due to long storage in the warehouse and therefore they must be cleaned. 


WARNING! Check that there are no water leaks in the pipe fittings. 


2)  Check that the suction tubes (located in the lower part of the machine) within the corresponding tanks containing the liquids provided (detergent, neutralizer, and/or others), are correctly inserted.

3)  Using the main circuit breaker, supply power to the machine.

4)  Check that the water supply flexible hoses are properly connected.

5)  At the beginning of the first cycle check the level probes of the liquid suction tubes (detergent, neutralizer and/or others): pull one tube at a time out of the container and check if the corresponding alarm flashes on the display, indicating the need for liquid refilling.

6)  After running 3-4 washing tests, clean the water filters placed in the lower compartment of the machine (Fig. 11).

7)  Check that the drain trap does not show water leaks and is firmly secured to both the machine and the drain.

8)  Check if the pumps correctly suck washing liquids. To do this, check that the liquid rises along the tube connected to them.

Loading the Chemicals

When changing the tanks chemicals, it is required to perform the manual filling of the hydraulic circuit. This phase is important because it prevents any air bubbles present in the pipes from causing an interruption of the cycle following an incorrect reading of the flow meters. 

To fill the pipes after changing or refilling the chemicals, proceed as follows: 

1) Open the door using the “DOOR” button on the touch screen (4), if the door is already open, leave it opened; 

2) Remove the rack if inserted, in order to have full visibility of the washing chamber; 

3)  Press and hold the button corresponding to the dosing pump you wish to activate to manually load the hydraulic system: key P1 (1) on the touch screen for pump DOS1; key P2 (2) on the touch screen for pump DOS2; key P3 (3) on the touch screen for pump DOS3; key P+ (4) on the touch screen for pump DOS4.

4)  Wait a few seconds to see in the lower right part, the area where the chemical inlet holes are present, a constant flow free from air bubbles.

5)  Release the button and repeat the sequence from 3 to 4 for the other dosing pumps that need to recharge the hydraulic circuit.


CAUTION! This phase must be performed at the first installation of the machine, in this case, let the liquid flow for a few more seconds to ensure the correct and full loading of the hydraulic system. 


Before Use

The glass-washer can be used for cleaning and rinsing glassware and laboratory utensils such as:

  • containers, test tubes, beakers, bottles and ampoules; – graduated containers: flasks, graduated cylinders; 
  • small glassware: lids, spatulas, magnetic laboratory stirrers, stoppers; 
  • other glassware such as petri dishes, funnels, pieces of tubing/cable, etc… 
Emergency Door Release

In the event of a blackout, or any other need in which it becomes difficult to open the glass-washer door, there is a manual emergency release that can only be activated if the door cannot be opened normally. 


HAZARD! If the emergency release is used while a program is running, very hot water and chemicals may leak. Therefore, there is a danger of burns, scalding and irritation. 


HAZARD! In the case of an emergency release of the door, the load might be contaminated. Be careful while handling it. 


Release the door:

1) Push the key supplied in the pack horizontally into the gap between the door and the lid or worktop (1 Fig. 14.1), to position the key correctly, follow the position indicated on the label (1 Fig. 14.2) on the door handle. 

2) Upon the door unlock take out the key. 

This is the first key slot graphic for the Tiva 8-L

This is the second key slot graphic for the Tiva 8-L

In case of program interruption, proceed with a new treatment of the load. 

Preparing the Load

WARNING! The maximum weight that the open door can withstand is 66 lbs (30 kg). The maximum volume of the wash chamber is 44 US gal (165 l). 


The glassware to be washed must be placed on the relative racks, making sure that they do not overlap one another. 


IMPORTANT NOTE! The cleaning quality depends on the correct loading of the glassware. 


Empty any residual liquids from the instruments and containers before inserting them in the chamber, washing all residues well (e.g. disinfectant solutions) in cold water.

Remove non-water soluble residues such as paint, adhesives and polymer compounds using appropriate solvents.

Scoop nutrient media (Agar) out of petri dishes.

Shake out any blood residues and remove any clots.

If necessary rinse the wash load briefly with water to avoid introducing coarse soiling into the machine.

Remove all stoppers, corks, labels, sealing wax residue, etc

Glassware wetted with solutions containing chloride or hydrochloric acid should first be rinsed in cold water and then placed in the drawers for washing. After use, they= should be placed on a settling tank for the time necessary to bring them to the rinse area. 


HAZARD! The amount of residual solvents and acids on items going into the chamber should be minimal. There should be no more than a trace of any solvents with a flash point of below 70°F (21°C). Chloride solutions, in particular hydrochloric acid, are corrosive for iron materials. 


Make sure that the glassware do not fall off the racks and inserts and do not hang from the grid bars.

Check that injector nozzles and connectors are securely held in position in the baskets or inserts.

Insert the glassware into the appropriate nozzles.

Position container-shaped objects so that liquids can flow easily and, if possible, place tall and heavy tools in the centre of the rack.

Place pointed or sharp objects in such a way as to prevent anyone from being injured by them.

Do not treat broken glass instruments with the glass-washer.


WARNING! Broken glass can result in serious injury when loading or unloading


Components with special geometries must be arranged so as to allow the water to flow out.

Introduce only suitable steel instruments, resistant to corrosion.

Objects made completely or partially of plastic must be resistant to high temperatures.

Nickel-plated and chrome-plated and/or aluminum instruments and tools may not always be suitable for machine treatment. They require special process conditions.

If possible, disassemble the modular instruments following the manufacturer’s instructions and treat the individual parts separately.

Treat small glassware and parts in special inserts.

Before the automatic treatment, check that the lumen and hollow body glassware can be treated internally and remove any obstructions.

Carefully place the load in the supports.

Objects must not be put inside each other nor cover each other. Do not place objects so close to each  other that they cannot be washed properly.

Arrange the load so that all surfaces can be reached by the washing liquid. Otherwise they will not be cleaned!

Place the objects so that the liquids can flow out without hindrance.

Tall, narrow, hollow and heavy items should be placed in the centre of the rack. This will ensure better water coverage. 


HAZARD! Be very careful when inserting pointed or sharp glassware. Always insert the glassware starting from the bottom, and possibly wearing protective gloves. 


When arranging the load, make sure that the spray arms are not blocked by the load itself.

After treatment, the transmission instruments must be cleaned according to the manufacturer’s instructions. After treatment, before reusing the transmission instruments, check that they are working properly, e.g. by spraying liquids into the sink.

Summary of the Rack Loading Operations

Depending on the load, special nozzles or adapters may be required for proper internal cleaning.

Sequence for loading the rack:

1) Fill the rack by arranging the glassware so that all surfaces are reached during washing. 

2)  Insert the rack into the chamber.

3)  Close the door and start the washing program. 

Wide-Necked Glassware

Wide-necked glassware (such as beakers, Erlenmeyer flasks and Petri dishes) or cylindrical glassware (such as test tubes) must be washed and rinsed both internally and externally. For this purpose, they should be placed in appropriate inserts and placed in an empty lower or upper rack.

Narrow-Necked Glassware

Narrow-necked glassware (such as narrow-necked Erlenmeyer flasks, flasks and graduated flasks) require the use of racks with special injection nozzles. 

Treatment of Glassware

WARNING! Insert only glassware suitable for automatic treatment in glasswasher, according to the manufacturer’s instructions. In particular, follow the information provided by the same manufacturer. Despite compliance with the manufacturer’s instructions, in case of damage or alteration of the glassware, the responsibility will be borne by the glassware manufacturer. 


Before starting the treatment, check that:

The external surface of the glassware is clean of material residues (E.g. chloride, hydrochloric acid etc).

The air and spray channels must be clean. 

Finally, carry out a test.

Use the recommended liquids (see “Using and Storing Chemicals” in our installation guide). 

Programs 

The machine leaves the factory with presetting washing programs already entered in the settings menu.

For programs details see the Washing Program booklet. 

Suggested Programs vs. Configuration 

For a good result, it’s important to choose the right program together with the right load/configuration. To select the programs available, use key P1, P2 or P3 depending on the most suitable program for the level of dirt of the load. 

Program Structure
  • Drain: It is used to empty the washing chamber.
  • Pre-wash: Pre-wash is necessary to eliminate coarse dirt and foamy substances.
  • Wash: Depending on the load, washing normally takes place at 113°F- 149°F (45°C- 65°C), adding the appropriate detergent. 
  • Rinse: The rinsing operation eliminates and neutralizes the chemicals of previous washes.
  • Drying: Sufficient drying reduces the risk of corrosion caused by the residual humidity on the load (not available for models without drying system). 

Starting the Machine

After checking the integrity and full efficiency of the machine, proceed with start-up:

Power the machine using the main circuit breaker.

Open the door (key 5) to introduce the racks. 

Before Starting the Program

Before starting each program, check:

That the filters, positioned on the bottom of the chamber (Fig. 11) are perfectly clean. Clean them, if necessary. 

The nozzles of the upper and lower spray arms must be free and clean.

Products must be arranged correctly.

The spray arms must be able to rotate freely. The machine constantly monitors the rotation speed during the program (optional).

Liquid containers must be sufficiently filled. Check for any messages on the display before starting the program.

Closing the Door 

Load the glassware and introduce the racks.

Close the door and push it until the lock is activated. The door can be unlocked and opened at any time before the program starts, by pressing the Door key (5)

Selecting the program

To select the programs available, use key P1, P2, P3 or the key P+ (to select programs from 4 to 40) on the control panel (shown above). 

Starting the Program


WARNING! Always follow the indicated procedures. Inattentive and frivolous use of electrical devices may lead to risks for the operator. 


The manufacturer is not liable for possible damage caused by uncontrolled use of the device.

After selecting the program with relative key, the display shows the selected program and the operating time and temperature. 

To start a program, press the START key (7, shown above). Time counter use the first cycle to self calculate the time for each cycle.

Program Execution

WARNING! The program can be interrupted at any time. However, it must be remembered that once interrupted, it must restart from the beginning. Only if it is interrupted in the drying phase can the cycle be considered completed successfully. The glassware must be dried.


HAZARD! In case of contact with the fluids inside the chamber after the cycle has been interrupted before the end of the rinse phase and consequent opening of the door, immediately contact the nearest health center.


Pre-Wash

Pre-wash is carried out with cold water (optional softened water) and without liquids. It mechanically dissolves encrusted organic materials and all proteins on the surface of the glassware. Avoid too high water temperatures. 

Wash

It is the actual cleaning cycle. The cleaning chamber heats up until it reaches the specific temperature for the selected program, a temperature that remains stable for the holding time. The cleaning liquid is automatically introduced before the start of the holding time. 

Rinsing

Rinsing is carried out with cold water, the machine is standard equipped with the second pump with which it is possible to carry out the neutralization cycle by introducing neutralizing liquid which is designed to reduce the alkalinity and clean the glassware from residues soluble in acid, e.g. limestone and rust. 

Dry

The glassware are dried internally and externally with filtered hot air. Good drying reduces the risk of corrosion caused by the residual humidity on the load. Glassware with a very small internal diameter must be dried further. 


HAZARD! If the drying phase is disabled, the machine will interrupt the cycle at the end of the rinse phase. It is the responsibility of the user to dry the glassware that have just been washed, using suitable cloths and following the instructions contained in the instruction manual for each individual glassware washed. 


HAZARD! If the drying phase is disabled, the glassware must NOT be left inside the machine, at the end of the cycle, wet, for longer than 5 minutes. This is to prevent the proliferation of bacteria and microorganisms. The manufacturer declines any liability for contamination of the glassware during the manual drying stage. 


Program End

The message «Finish» with a green background on the display indicates that the program has been carried out correctly. Unlock the door by pressing the Door key (5 in the panel picture from “Programs” section) and open it. 


IMPORTANT NOTE! Open the door immediately after the end of the program to avoid condensation forming. 


Check the results at the end of the cleaning process. The glassware must be completely clean and dry. 


IMPORTANT NOTE! it is the responsibility of the user to ensure that all instruments are dry. If the instruments are still wet: check that the correct washing program has been selected according to the load configuration and that the drying filters have been regularly checked. 


  • Check hollow glass ware and retreat them if necessary. The hole (lumen) of hollow glassware must be free.
  • The injectors must be correctly placed on the connection tube in the cleaning chamber.
  • The nozzles and connections to the base grid must be firmly fixed.

If these checks are positive and the program has been carried out without interruptions or malfunctions, the load has been successfully cleaned.

Extracting the Load 

At the end of the program and when extracting the load:

Do not force the door open to avoid damaging the device or creating a hazardous condition. 


WARNING! Particularly large glassware can be very hot at the end of the program. Allow  glassware to cool down before removing them. Use adequate gloves that protects against burns. 


HAZARD! Failure to follow recommendations can cause burns. 


 

Control Panel

The control panel consists of 8 keys and a 3.5 inch LCD display. All keys, except key “3”, are multi-function, depending on the action being performed in a specified state of the device. 

Here below the description of the control panel standard (STD) and special (SPC) functions. Later on, with the description of how the machine works, we will find these symbols again. 


IMPORTANT NOTE! USER PIN MANAGEMENT option on request. The special functions (SPC) related to the username and password settings for the USER are not available if the option is not enable. 


1.) P1 

  • STD: select program 1;
  • SPC: “1” character for the custom password; 
  • SPC: navigate upward in the menu;
  • SPC: change the value of the highlighted parameter. 

2.) P2 

  • STD: select program 2;
  • SPC: “2” character for the custom password; 
  • SPC: navigate forward in the menu;
  • SPC: change the value of the highlighted parameter. 

3.) P3 

  • STD: select program 3; 
  • SPC: “3” character for the custom password; 

4.) P+ 

  • STD: select the next program (proceeding by +1 at a time upon program 40);
  • SPC: delete incorrectly typed field. 

5.) Door 

  • STD: open the door at the end of the program; 
  • SPC: exit from parameters and programs. 

6.) Prg 

  • STD: enter to the program menu; 
  • SPC: “C” character for the custom password. 

7.) Start 

  • STD: enter the selected menu item;
  • SPC: continue to the next parameter;
  • SPC: “B” character for the custom password. 

8.) Reset 

  • STD: stop a running program;
  • SPC: return to the previous menu item;
  • SPC: “A” character for the custom password. 
Colors of the Keys

Depending on the menu function you are using, the keys get highlighted with different colors:

  • White light: indicates the keys that can be used in that particular action. 
  • Red light: indicates the function “go back to the previous menu” if associated with the Door key; indicates the function “delete the characters typed” if associated with the P+ key.
  • Green light: indicates confirmation of the action. 

IMPORTANT NOTE! On any screen and for any action you can only use the highlighted keys.


Display

The LCD display shows the status of the machine. The images show the various phases and the current operation. For example: if the machine is loading water, the display will show the image of the tap, of the water flow meter and of the water being loaded. If water is cold, droplets will be blue and if water is hot, they will be red. The temperatures are set in degrees Celsius or Fahrenheit. The images are animated, to show how the current operation evolves. 

This is the description of the display graphic for the Tiva 8-L.
Description of the Display (shown above)

1) The number of the running program is displayed (e.g. P02 = program 2);

2) It displays the phase the machine is running. If the machine is going to run a wash cycle it will display the writing «Wash». When the machine displays an alarm that space background turns red and the message shows the number of the alarm and a brief description; 

3) Water conductivity value in microsiemens (if conductivity probe function is enabled); 

4) Several animated images showing the current status of the machine are displayed; 

5) The elapsed phase time from the moment the temperature set for the phase (see 12) has been reached is displayed; 

6) Bar indicating the progress of the program; if the program is ending, the bar will be almost entirely green; 

7) The temperature measured by the PT1000 probe, placed after the air heater, is displayed to indicate the temperature of the air entering the chamber; 

8) It displays the temperature measured by the first PT1000 probe placed in the chamber; 

9) It is the display of the temperature measured by the second PT1000 probe placed in the chamber. The temperature detectable by the two probes must not differ from each other by more than 36°F (2°C); 

10) Time for which the device should maintain the set temperature (see 12); 

11) Phase of the program;

12) Temperature set for the current phase;

13) Date and time are displayed with the machine in stand-by. 

Messages on the Display 

On the display of rack insertion, messages may be displayed, such as:

  • Product 1 liquid reserve: meaning that the liquid inside the product 1 tank is finishing and must be replaced;
  • Product 2 liquid reserve: meaning that the liquid inside the product 2 tank is finishing and must be replaced.

When the cycle has finished the following message appears: «Finish» and a green light appears in the chamber. At this point the door is unlocked and you can unload the rack with the washed items. 

Bluetooth Enabled

If bluetooth is enabled, the following icon will appear on some display screens: 

This is a bluetooth connected graphic.

If bluetooth has been disabled the following icon will be displayed: 

This is a bluetooth disconnected graphic.

For more information on bluetooth, see chapter “Bluetooth”. 

Display Screens 

After starting the machine, following the instructions in chapter “Starting the machine”, the display will show the images indicating step-by-step the operations in progress. 

1.) Once the machine has started, the display shows the screen indicating door open and the invitation to insert the rack with the instruments to be washed. Then insert the rack with the instruments and close the door. 


IMPORTANT NOTE! The door must be tightly closed until you hear the classic closing «click», otherwise the program will not start. 


This graphic shows to insert basket and close the door on the Tiva 8-L.2.) With the machine started and door closed, the screen for selecting the programs will appear. Press the selected key (P1 or P2 or P3 Fig. 13) on the control panel. To access subsequent programs (if stored), press the key «P+» several times. 

This is the select program graphic from the Tiva 8-L.

3.) Optional: once the program is selected, the screen for selecting the user will appear. Press the key P1 or P2 on the control panel to navigate upward and forward in the user list. Then press Start.

4.) Optional: enter the user password: use the key P1, P2, P3, Reset, Start and Prg on the control panel to digit the correct password (See Fig. 13 for key description). 

This is the user password graphic from the Tiva 8-L.

Program 2 Standard

Phase 1 

The machine starts the automatic work cycle and discharges any residual water present in the chamber. 

This is the standard drain graphic from the Tiva 8-L from the beginning of the cycle.Phase 2

Self-loading of cold water. During the self-loading phase the chamber will be filled. 

This is the standard fill cold water graphic from the Tiva 8-L.Prewash phase begins. 

This is the standard pre-wash graphic from the Tiva 8-L.Phase 3

At the end of the prewash phase, the water is automatically discharged. 

This is the standard drain graphic from the Tiva 8-L from the beginning of the cycle.Phase 4

Self-loading of hot water. During the self-loading phase the chamber will be filled. 

This is the standard fill hot water graphic from the Tiva 8-L.Wash phase begins: the machine will rise the water temperature to the preset value and will keep it for the preset time. 

This is the standard wash graphic from the Tiva 8-L.

When the water reaches the 95°F (35°C), the peristaltic pump 1 will dose the detergent. 

This is the screen displayed when the first product enters the Tiva 8-L cycle. Phase 5

At the end of the wash phase, the water is automatically discharged. 

This is the standard drain graphic from the Tiva 8-L from the end of the first wash.

Phase 6

Self-loading of hot water. During the self-loading phase the chamber will be filled. 

This is the standard fill hot water graphic from the second self-load of the Tiva 8-L cycle.Rinse phase begins: the machine will rise the water temperature to the preset value and will keep it for the preset time. 

This is the standard rinse graphic from the Tiva 8-L.When the water reaches the 35°C (95°F), the peristaltic pump 2 will dose the neutralizing agent. 

This is the screen displayed when the second product enters the Tiva 8-L cycle. Phase 7

At the end of the rinse phase, the water is automatically discharged. 

Phase 8

Self-loading of deionized water. During the self- loading phase the chamber will be filled. 

Rinse phase begins: the machine will rinse the instruments while checking the conductivity of the water. At the end of the phase, if the conductivity will be below a preset value, the machine will proceed to the next phase. If the check is negative, the machine will repeat phases 7 and 8 again. If the second check will fail, the machine will show an error and stop the cycle. 

Phase 9

At the end of the rinse phase, the water is automatically discharged. 

Phase 10 

Dry phase begins: hot air is blown inside the chamber at a preset temperature and kept for the preset time. Phase not present in the version without drying. 

End of the «STANDARD» cycle. Open the door and remove the rack. 

If, for any reason, it is necessary to interrupt the cycle, simply hold down the RESET key for a few seconds until an audible alarm (buzzer) is heard and an alarm screen appears on the display. 

Once the problem is resolved, resume the cycle from the beginning. If the problem cannot be solved, contact the technical assistance. 

Alarm Messages 

The machine is equipped with an alarm system that indicates malfunctions detectable with an audible signal (buzzer), a screen on the graphic display of the control panel and a red led light in the camber. 

At first the image concerning the alarm is displayed (for 5 seconds); then the alarm description is displayed (for 10 seconds).
The image and text alternate until the alarm is reset. 

If the program is interrupted because of an alarm, the display will show the message «Program interrupted stop-alarm». Follow the message shown on the display to reset the alarm. 


IMPORTANT NOTE! If the RESET button is pressed during the wash cycle, the program is interrupted and an alarm is displayed with the message «Program interrupted stop – user». Press RESET to return to initial display. 


For the alarms, description and possible solutions see Annex 12.6. 

User Settings Menu 

Start screen 

To access the settings menu the procedure is as follows: Open door -> PRG key for 5 seconds-> If the user and password are enabled, insert user password. 


IMPORTANT NOTE! USER PIN MANAGEMENT option on request. 


The keys and their operation are shown below, contextualized according the menu in which you are located:

  • Use P1 and P2 to navigate up and down the menu.
  • Press START to access the entry of the menu selected. 
  • Within the parameter, use the 1 and 2 keys to change the value of the highlighted parameter.
  • Press START to go to the next parameter.
  • Press PRG to return to the previous menu. 

The menu consists of a screen, the item, when selected, lights up: 

Settings

See Annexes 12.1 and 12.2 below.

Annex 12.1

Annex 12.2

Language Setup

To change the display language, with the door open, press PRG for 5 seconds and enter the Super User password. Once the menu appears, press the START button under GENERAL. Scroll to the “Display Language” setting and select the desired language from those available. 

USB key Management

When the USB key is hooked up in the machine, the following message appears after a few seconds: 

When the USB key is removed from the machine, after a few seconds the following message appears:

Use USB 2.0 only, system does not support USB 3.0. 

Procedure for Password Change

To generate the password use the keys P1, P2, P3, Reset, Start and Prg (See panel picture from before for key description). Entering the password incorrectly three times in a row will cause the block of the user. 

Annex 12.3

This is the Tuttnauer Tiva 8-L Annex 12.3.

Expiring Passwords

User Password

Has a preset validity of 6 months, after the expiring date, it must be renewed. At first access it must be generated by the Super User. See Annex 12.3 for the procedure for password change. See Annex 12.4 below for the procedure for user password setup for the Super User. This option is available if the USER PIN MANAGEMENT is enable. 

Annex 12.4

This is the Tuttnauer Tiva 8-L Annex 12.4.Super User Password

At the first access enter the password “111111″. As for the User, the Super User password has a preset validity of 6 months, after the expiring date, it must be renewed. See Annex 12.5 below.

Annex 12.5 

This is the Tuttnauer Tiva 8-L Annex 12.5.Alarms
This is the Tuttnauer Tiva 8-L Annex 12.6 page 1.
This is the Tuttnauer Tiva 8-L Annex 12.6 page 2.
This is the Tuttnauer Tiva 8-L Annex 12.6 page 3.Warnings

This is the Tuttnauer Tiva 8-L Annex 12.7.


As always if you have any questions about this process or anything else please feel free to contact us and take advantage of our “FREE TECH SUPPORT.”

We also offerFREE VIRTUAL TECH SUPPORTto “See and Talk” with a “Real Time Live Technician” for any problems you may be in need of help with.

You can also use our “FREE MAINTENANCE PROGRAM”. Take the guesswork and worrying about what unit is due for maintenance and which maintenance cycle it is time for. We will keep track of all your autoclaves and let you know when it’s time for anything.

SHOPPING CART

close