This is the Tuttnauer T-Top Maintenance Guide.

Tuttnauer T-Top Maintenance Guide

The Tuttnauer T-Top autoclave is Tuttnauer’s newest machine that we have the pleasure to bring to SAS. This 21-liter mechanical marvel was designed for small to medium This is a Tuttnauer T-Top Autoclave 21L Tabletop Sterilizer medical and dental practices. Aiming to maximize workflow through a wealth of exciting modern features. Recently, we’ve written about the Tuttnauer T-Top, going in-depth on it’s features and specs, and created an installation guide using manufacturer’s instructions as well here. 

But in this article, we’ll be going in-depth on the operation process for the Tuttnauer T-Top. This Tuttnauer T-Top operation guide comes directly from Tuttnauer. Ensuring that the installation process is done properly to manufacturer specifications.

Maintenance Guide

Warnings! 

  • Only technical personnel having proper qualifications and holding technical documentation (including a technician manual) and adequate information are authorized to install and serve the apparatus. 
  • Make sure the autoclave is not hot before cleaning it. Before proceeding, make sure that the electric power is disconnected and there is no pressure in the chamber. 
  • Perform maintenance operations as instructed.

The maintenance operations described in this chapter need to be followed as indicated to keep the device in good working condition and to keep any breakdown time to a minimum. 

Should the need arise, technical assistance or a service technician can be requested by either calling your dealer or Tuttnauer Europe . 

Daily Maintenance

Caution!

Make sure the autoclave is not hot before cleaning it. 

  • Turn the unit on momentarily to allow the door to be opened. Open the door, unplug the autoclave again, and proceed with cleaning. 
  • Clean door gasket with a mild detergent, water and a soft cloth or sponge. Check visually that the gasket is intact, not loose and clean. 

Weekly Maintenance

Caution!

Make sure the autoclave is not hot before cleaning it. 

  • Turn the unit on momentarily to allow the door to be opened. Open the door, unplug the autoclave again, and proceed with cleaning. 
  • If the autoclave is only used occasionally, drain the water from the mineral free water reservoir once a week, and refill with fresh mineral-free water or distilled water. 
  • Once a week or if a text alarm of ’full waste water reservoir’ appears (whichever comes first) drain the water from the waste water reservoir. 
  • Clean the outer parts of the autoclave with a soft cloth. 
System Clean

Once every two weeks perform system clean process per section 11.

Monthly Maintenance

Caution!

Make sure the autoclave is not hot before cleaning it. 

  • Turn the unit on momentarily to allow the door to be opened. Open the door, unplug the autoclave again, and proceed with cleaning. 
  • Clean and descale the chamber.
  • Clean the outer parts of the autoclave with a soft cloth.
  • Clean the Drain filter of the autoclave. 

Periodic Maintenance

Caution! Make sure the autoclave is not hot before cleaning it 

  1. Replace the air filter, every 6 months or after 1000 cycles (whichever comes first). 
  2. Every 3 months check the door gasket for any signs of physical damage and ask the technician to replace it if there is tear or leakage. 

Replacing the Air Filter

A caution symbolBefore proceeding, make sure that the electric cord is disconnected and there is no pressure in the chamber. Use scissors to open the filter bag and not sharp blades or pointed instrument. Check that the new filter has not exceeded the maximum shelf life. Carefully un-pack the new filter and examine it for any signs of damage. Remove any protective packaging before inserting the filter into place. 

The AIR filter is located, when opening the door, in the upper left corner . 

  1. This is the location of the Tuttnauer T-Top Air Filter.Pull out the filter. 
  2. Connect the new filter.
  3. Ensure the new filter is all the way in and seated properly. 

Note: Make sure that the arrow on the filter body points inwards, toward the chamber. Make sure that you don’t bend the filter pipe when reattaching it. 

Cleaning the Drain Filter

Make sure the autoclave is not hot before cleaning it 

Before proceeding, make sure that the electric cord is disconnected and there is no pressure in the chamber. 

1.) Clean the drain filter every month. 

The drain filter is located inside the autoclave chamber at the bottom far end. To reach the filter, open the chamber door and remove the tray holder. 

2.) Open the autoclave chamber door and remove the tray. 

This is the open door and tray layout of the Tuttnauer T-Top.3.) Clean with a soft cloth the filter and the area around the filter. 

This is the location of the Tuttnauer T-Top drain filter.4.) Check that the autoclave functions normally after the drain filter cleaning. 

Note: If by cleaning the filter the result is not satisfactory, proceed to the replacement of the drain filter. (paragraph 8.4.3 ). 

Replacing the Drain Filter

Note: If after successive cleaning of the filter the result is not satisfactory proceed to the replacement of the drain filter. 

Before proceeding, make sure that the electric cord is disconnected and there is no pressure in the chamber. 

Use scissors to open the filter bag and not sharp blades or pointed instrument. Carefully un-pack the new filter and examine it for any signs of damage. Remove any protective packaging before inserting the filter into place. 

Caution! Make sure the autoclave is not hot before cleaning it 

The drain filter is located inside the autoclave chamber at the bottom far end. To reach the filter, open the chamber door and remove the tray holder. 

1.) Open the autoclave chamber door and remove the tray. 

2.) Unscrew the old Drain filter assembly (screw+net). 

This is the Tuttnauer T-Top drain filter assembly in the chamber.3.) Screw the new Drain filter assembly, (screw + net) to its place in the chamber bottom, tighten it manually. 

This is the disassembled Tuttnauer T-Top drain valve assembly. 4.) Place the tray back.


As always if you have any questions about this process or anything else please feel free to contact us and take advantage of our “FREE TECH SUPPORT.”

We also offer FREE VIRTUAL TECH SUPPORT to “See and Talk” with a “Real Time Live Technician” for any problems you may be in need of help with.

You can also use our “FREE MAINTENANCE PROGRAM”. Take the guesswork and worrying about what unit is due for maintenance and which maintenance cycle it is time for. We will keep track of all your autoclaves and let you know when it’s time for anything.


You can find the Tuttnauer T-Top autoclave on our website here.

SHOPPING CART

close