Tuttnauer T-Top Installation Guide

The Tuttnauer T-Top autoclave is Tuttnauer’s newest machine that we have the pleasure to bring to SAS. This 21-liter mechanical marvel was designed for small to mediumThis is a Tuttnauer T-Top Autoclave 21L Tabletop Sterilizer medical and dental practices. Aiming to maximize workflow through a wealth of exciting modern features. Recently, we’ve written about the Tuttnauer T-Top, going in-depth on it’s features and specs, which you can find here on our site. And we also have the Tuttnauer T-Top available for purchase on our site here. 

But in this article, we’ll be going in-depth on the installation process for the Tuttnauer T-Top. This Tuttnauer T-Top installation guide comes directly from Tuttnauer. Ensuring that the installation process is done properly to manufacturer specifications.

Installation Guide

Lifting and Carrying

A caution symbolWarnings!

  • Only technical personnel having proper qualifications and holding technical documentation (including a Technician Manual), and adequate information are authorized to install and serve the apparatus.
  • Before moving the autoclave, make sure that the electric cord is disconnected from the power, and there is no pressure in the chamber.
  • Mind the power socket.Keep it and its vicinity dry.Danger of electrocution.
  • Drain the water from the reservoir (see section Draining the Reservoirs).
  • Do not drop the device!
  • To avoid injuries, lifting and carrying should be done with at least two persons or by using a fork-lift or any other mechanical aid.
  • For devices that weigh less than 75 Kg: The device is not designed for use on any standard slide out shelf. If it is necessary to use a slide out shelf, it must be tested and/or rated for 75 Kg or more.
  • Insufficient space for ventilation may result in malfunction or damage due to overheating.
Device Placement and Operating Conditions
  1. The autoclave is intended for indoor use only 
  2. Check and verify that the counter carrying the autoclave is a rigid and leveled surface and can carry a load of 75 kg. 
  3. Check and verify that the dimensions of the surface of the counter are at least 60 cm x 53 cm
  4. Keep the back and the sides of the autoclave approximately 10 mm away from the wall to allow ventilation and facilitate the device disconnection.
  5. If placed in a cabinet, verify that the rear of the cabinet is open to allow ventilation
  6. Keep a 50 cm space above the autoclave for filling water in the reservoir
  7. It is recommended that enough space be left around the autoclave to give a technician access for servicing the machine. 
  8. Check and verify that the ambient temperature range is 5°C-40°C. 
  9. Check and verify that the ambient relative humidity does not exceed 80%.
  10. The operational altitude shall not be over 13123ft (4000meters).
  11. Ambient pressure shall not be lower than 60.5 KPa.
  12. Operate the autoclave only in the manner specified in the manual. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. 
Connections to Utility Supplies
  1. Check and verify that the power supply is a 1 phase, 230 VAC ±10%, 50Hz. 
  2. Check and verify that the electrical net is protected by a current leakage safety relay. 
  3. The autoclave is defined as a Class A device, marketed for use in a commercial, industrial, or business environment. 
Storage

After the removal of the autoclave from the package, we recommend the following: 

  • Keep the device dry.
  • Keep the device away from sunlight and protect it from heat. 
Initial Operation of the Device

Note: Remove all external packaging material before turning ON the device. 

1.) Plug the power cord into the power socket. 

This is the Tuttnauer T-Top power socket location graphic.Warning! Mind the power socket. Keep it and its vicinity dry. Danger of electrocution. 

3.) Turn the ON/OFF Switch mounted on the front panel, behind the control door. 

This is the Tuttnauer T-Top power switch location graphic.4.) When you turn on the autoclave, it will automatically warm up.

Warning! Be careful the surfaces may be hot!

5.) Fill the Mineral Free Water Reservoir with water meeting the quality specs (see Water Quality and Filling the Mineral-Free Water Reservoir).

6.) Set date and time (see Set date and time).

Before Starting the Autoclave

Note: The device automatically resets the Atmospheric Pressure when turned on.
Open the Autoclave door for 2 minutes and verify that the ambient temperature is below 45°C. 

To reset the Atmospheric Pressure manually: 

1.) In the Quick Option screen, press the Settings icon (the gear icon) then browse to Maintenance/Reset atmospheric pressure. 

2.) In the confirmation screen, press “Reset.”

This is the Tuttnauer T-Top atmospheric pressure reset button confirmation screen.Wi-Fi Configuration 

Prior to operating the autoclave for the first time, please connect the Wi-Fi and ensure it remains connected at all times. 

The Wi-Fi connection is used for uploading all data information from devices in the field, to the online company’s server. 

This data is used for: 

  • Monitoring 
  • Viewing and downloading device history 
  • Software updates while validating the integrity of the update (applicable for users with Tuttnauer code) 

The data is backed-up on the device main board. 

Note: In case of connection failure, the data may be exported manually. 

To connect the Wi-Fi: 

1.) In the Quick Option screen, press the Wifi Configuration option.

2.) To enable the connection, press “On.”

3.) The Wifi Configuration screen is displayed with the available networks. 

This is the Tuttnauer T-Top WiFi configuration instructional graphic. 4.) Select your network, enter your Password and press the Accept privacy policy. 

This graphic is the Tuttnauer T-Top WiFi password entry screen. 5.) Press “Connect.”

6.) The Wi-FI is successfully connected. Press “Ok.” 

Checking the Device Version information.

1.) In the Quick Option screen, press the Info option (the “i” in the circle).

2.) Select the Version information.

This is the Tuttnauer T-Top version information button location graphic. 3.) The Version information is displayed/ For an example, see below:

This is the Tuttnauer T-Top version information screen.4.) Press on the backward arrow to return to the Info screen.


As always if you have any questions about this process or anything else please feel free to contact us and take advantage of our “FREE TECH SUPPORT.”

We also offer FREE VIRTUAL TECH SUPPORT to “See and Talk” with a “Real Time Live Technician” for any problems you may be in need of help with.

You can also use our “FREE MAINTENANCE PROGRAM”. Take the guesswork and worrying about what unit is due for maintenance and which maintenance cycle it is time for. We will keep track of all your autoclaves and let you know when it’s time for anything.


You can find the Tuttnauer T-Top autoclave on our website here.