This is the SteriDent Static Heat Sterilizer Model #200 Maintenance Guide.

SteriDent Static Heat Sterilizer Model #200 Maintenance Guide

The SteriDent Static Heat Sterilizer Model #200 is here at SAS, and in this article we’re going to cover the unit’s Maintenance Guide. We have a previously writtenThis is a SteriDent Static Heat Sterilizer Model #200 - #300 120 VAC 301000-SD Operation Guide that you can find here on our site for proper operation of your unit. As well as an overview on the SteriDent Model #200 itself. The SteriDent Static Heat Sterilizer Model #200 is a cutting-edge autoclave designed to meet the sterilization needs of dental and medical practices. Featuring a compact and sleek design, and offers ease of use and efficient performance.

And this article will be dedicated to the maintenance instructions provided by the manufacturer. Proper maintenance of any autoclave is important to ensure that nothing wears down or fails, along with extending the overall longevity of your unit.

 


Maintenance Guide

 

CLEANING AND MAINTENANCE

The Steri-DENT sterilizer is designed for minimum and easy service. Clean the inside and outside of the sterilizer with a stainless steel cleaner at least once a month. We recommend Cleaning Kit #201009. When cleaning the inside of the chamber, spray cleaner on cloth first and avoid contact with heating elements. Use a wet cloth to wipe away any cleaner residue and then follow with a dry cloth to wipe dry.

All of the electrical and mechanical components used in the sterilizer’s construction are long life, heavy-duty parts that require no maintenance. An authorized service representative or CPAC should be contacted if any of these parts need to be replaced.

Replace the inner door gasket once a year, or sooner if it becomes loose or worn. Replace the thermostat if the temperature control becomes erratic or every 3 years. 

ENVIRONMENTAL CONDITIONS
  • The Steri-Dent sterilizer is designed for indoor use with the following conditions:
  • Temperature Range of 5°C to 40°C (41°F to 104°F)
  • Maximum Relative Humidity of 80% up to 31°C (88°F).

Decreasing linearly to 50% at 40°C (104°F).

  • Pollution Degree 2 applies in accordance with IE 664.
  • Transient Over-voltage Category II applies.
  • Supply voltage not to fluctuate more than 10% (+/-12V at 120V, +/-22V at 220V)
  • Maximum altitude of 2000 m (6562 ft).

If you have any questions about the SteriDent Static Heat Sterilizer Model #200, any other CPAC units or autoclaves in general, please give us a call at 704-966-1650 option 3 for our Free Tech Support line. Or you can find our various programs using the links below.


As always if you have any questions about this process or anything else please feel free to contact us and take advantage of our “FREE TECH SUPPORT.”