In this article, we’re going to explain one of the many error codes that can affect your Midmark M9/Midmark M11 autoclave. This article will focus specifically on error code C671. We’ll tell you about the problem, what displays when the error occurs, its possible causes, and the course of action to take to resolve said error.
However, if you’re more of a visual learner, we have a video on this error code available here.
And if you just want to talk to someone, call our Free Tech Support at 704-966-1650 Option 3.
Display/Symptom
C671: SELECT MODE PRESSURE OVERLIM
UNPLUG / REPLUG UNIT
Gauge pressure inside chamber is greater than 34.8 psig (240 kPa) during SELECT mode.
Causes, Checks and Actions
Possible Cause 1: PC Board malfunctioning.
- Check 1: Perform: Heating Element Supply Voltage Test (With harness attached to leads should have line voltage in test mode)
- Check 2: Perform: Main PC Board Relay Test (Check TP1 & TP2 for an acceptable range: 10 to 14 VDC)
- Check 3: Perform: Main PC Board Pressure Transducer Voltage Test (Acceptable range: 4.0 to 6.0 VDC)
- Action: Replace Main PC board.