In this article, we’re going to explain one of the many error codes that can affect your Midmark M9/Midmark M11 autoclave. This article will focus specifically on error code C384. We’ll tell you about the problem, what displays when the error occurs, its possible causes, and the course of action to take to resolve said error.
However if you’re more of a visual learner we have a video on this error code available here.
And if you just want to talk to someone, call our Free Tech Support at 704-966-1650 Option 3.
Display/Symptom
C384: STERILIZE MODE DOOR OPEN
ITEMS NOT STERILE UNPLUG / REPLUG UNIT
Causes, Checks and Actions
Possible Cause 1: Door Switch/PC Board malfunctioning.
Note: Test switch with the door OPEN and Closed.
Check 1: Perform: Door Switch Continuity Test. With door closed, disconnect door switch wires from J1 & J2 of Main PC board. Place meter probes on door switch wires. If Continuity is not present replace door switch. Refer to: Door Switch Continuity Test
Action 1: Replace door switch.
Midmark Door Switch available here.
- Check 2: Check supply voltage from PC board. (Line voltage 120 or 230 VAC) Refer to: Door Switch Supply Voltage Test
- Action 2: Replace main PC board.
As always if you have any questions about this process or anything else please feel free to contact us and take advantage of our “FREE TECH SUPPORT.”