This article explains the possible causes and solutions to a Midmark M9-M11 New Style autoclave Error Code C060.

Midmark M9/M11 Error Code C060

In this article, we’re going to explain one of the many error codes that can affect your Midmark M9/Midmark M11 autoclave. This article will focus specifically on error code C060. We’ll tell you about the problem, what displays when the error occurs, its possible causes, and the course of action to take to resolve said error.

However if you’re more of a visual learner we have a video on this error code available here. 

And if you just want to talk to someone, call our Free Tech Support at 704-966-1650 Option 3.

 

Display/Symptom

C060: POWER MODE SYSTEM HARDWARE

ITEMS NOT STERILE

UNPLUG/RE-PLUG UNIT

Causes, Checks and Actions

Possible Cause 1: Power interruption. (Unit lost power for a few seconds during Fill, Heat-Up, or Sterilization Mode.)
  • Check 1: Check power cord connections. (Power interruption could be due to an electrical storm, brown out, etc.)
  • Action 1:Unplug unit for 60 seconds, then plug unit into a dedicated, properly rated circuit.

 

  • Check 2: Check supply voltage.
  • Action 2: Contact an electrician to modify voltage.

Possible Cause 2: Loose / Damaged wire connections.
  • Check: Check AC terminals and connections at J5 on main PC board
  • Action: Secure / Repair loose or damaged connections.

Possible Cause 3: Power cord/receptacle damaged.
  • Check: Inspect condition of power cord/ receptacle.
  • Action: Replace power cord/ receptacle.
Midmark M9/M11 Power Cord Receptacle is available here. 

Possible Cause 4: Ribbon cable loose or damaged.
  • Check: If unit is beeping/clicking with no display. (or display cuts out)
  • Action: Secure/replace ribbon harness.
Midmark Ribbon Cable available here. 

Possible Cause 5: Display PC compromised.
  • Check: Check for damaged pins/connection to ribbon cable.
  • Action: Replace display PC board.
Midmark Display PC Board is available here. 

Possible Cause 6: Error Code will not reset.
  • Check: Unit must be unplugged for 60 seconds.
  • Action: Replace Main PC board.
Midmark M9 Main PC Board available to buy here. 

As always if you have any questions about this process or anything else please feel free to contact us and take advantage of our “FREE TECH SUPPORT.”

We also offer FREE VIRTUAL TECH SUPPORT to “See and Talk” with a “Real Time Live Technician” for any problems you may be in need of help with.

You can also use our “FREE MAINTENANCE PROGRAM”. Take the guesswork and worrying about what unit is due for maintenance and which maintenance cycle it is time for. We will keep track of all your autoclaves and let you know when it’s time for anything.


The Midmark M9 autoclave is available here. 

We also have the Midmark M11 autoclave here. 