This is the Solmetex DryShield Autoclavable Mouthpiece Maintenance Guide

Solmetex DryShield Autoclavable Mouthpiece Maintenance Guide

We’ve gone over the ins and outs of Solmetex’s DryShield Mouthpieces in detail before, including the installation guides for both their Autoclavable and Single-Use mouthpieceThis is the Medium DryShield autoclavable mouthpiece DS-MPM-004 options. Which means it’s now time to cover a step exclusive to their Autoclavable mouthpieces, cleaning and maintenance.

So in this article, we’ll be covering the Solmetex DryShield Autoclavable Mouthpiece Maintenance Guide. These are instructions directly from Solmetex that detail proper cleaning and maintenance procedures for your autoclavable mouthpieces. And if you have any questions about these instructions, Solmetex Mouthpieces or anything else, give us a call at 704-966-1650.

 


Maintenance Guide

Cleaning, Sterilization and Maintenance

Things You Should Know

The DryShield DS1 is made from a high heat-tolerant material, and the DryShield mouthpiece is made of an autoclavable silicone material. All DryShield components are completely autoclavable. DryShield recommends following standard infection control protocols for sterilization.

DryShield DS1

  • After completion of dental procedure, DryShield recommends running the DryShield DS1 with a non-corrosive, pH neutral enzymatic cleaning solution and performing a surface disinfection with a disinfectant wipe between each patient. Users have the option to sterilize the DryShield DS1.
  • Maintenance of the lever o-ring is done through visual inspection. Replace o-ring when lever function decreases.

DryShield Autoclavable Mouthpiece

  • After completion of each dental procedure, detach mouthpiece from bite block and place both in ultrasonic cleaner for 15 minutes.
  • Next, clean the separated mouthpiece and bite block of debris by thoroughly scrubbing with soap and water or a detergent solution. Clean the inside of the mouthpiece through access at the slit opening and at the suction connector region using a long-handled brush and thorough scrubbing.
  • Then, rinse under running water, air dry and place in autoclave pouch to be sterilized.
  • The DryShield Mouthpiece MUST BE sterilized between each use. Sterilize in a steam autoclave machine at minimum 121 °C (250 °F) for a minimum of 30 minutes. Allow a minimum of 30 minutes in drying cycle.
  • The sterilization parameters for the DryShield Mouthpiece are found below.
Cycle Type ParameterMinimum Set Point
Gravity 250° F (121°C)Exposure Temperature250° F (121°C)
Exposure Time30 minutes
Dry Time 30 minutes

As always if you have any questions about this process or anything else please feel free to contact us and take advantage of our “FREE TECH SUPPORT.”