This graphic explains the proper cleaning procedure for the chamber and plumbing like this:
A) With a cooled chamber, drain the sterilizer’s reservoir and refill with clean distilled water or water that meets the referenced water purity specifications. Add one ounce of Speed-Clean sterilizer cleaner directly to the bottom of chamber.
B) Run one Pouches cycle.
C) Press Stop button when Dry Cycle begins. (Dry Cycle is not needed during maintenance.)
D) Drain reservoir and refill a second time with clean distilled water or water that meets the referenced water purity specifications.
E) Rinse by running one Unwrapped cycle . Push the “Stop” button when the drying cycle begins.
F) Drain and refill reservoir with clean distilled water or water that meets the referenced water purity specifications, then allow sterilizer to cool.
G) Remove tray plate, rack, and trays. Wipe off with a damp cloth.
