This is the Tuttnauer T-Edge 11S Installation Guide.

Tuttnauer T-Edge 11S Installation Guide

In this article, we’ll be going in-depth on the installation process for the Tuttnauer T-Edge 11S. We’ve written an overview about the T-Edge 11S which you can findThis graphic shows the Tuttnauer T-Edge 11S autoclave. here on our site. The T-Edge 11S is a modern autoclave with lots of benefits for your practice, and even a unique feature. The T-Edge 11S has the ability to be upgraded from a gravity displacement cycle to a pre/post vacuum cycle. With all that and more, the T-Edge 11S can be a great addition to any practice.

This Tuttnauer T-Edge 11S installation guide comes directly from Tuttnauer. Ensuring that the installation process is done properly to manufacturer specifications. If you’d like to see a digital version of the T-Edge 11S manual, we have a link to it on our website from our resources tab here.


Installation Guide

Lifting and Carrying

A caution symbol Cautions! 

Before moving the autoclave, make sure that the electric cord is disconnected from the power, and there is no pressure in the chamber. 

Drain the water from the reservoir

Do not drop the device! 

To avoid injuries, lifting and carrying should be done with at least two people or by using a fork-lift or any other mechanical aid. 


Device Placement and Operating Conditions 

1.) The autoclave is intended for indoor use only. 

2.) Check and verify that the counter carrying the autoclave is a rigid and leveled surface and can carry a load of 165 lbs (75kg). 

A caution symbol Caution! The device is not designed for use on any standard slide out shelf. If it is necessary to use a slide out shelf, it must be tested and/or rated for 165 lbs (75Kg) or more. 

3.) Check and verify that the dimensions of the surface of the counter are at least 21″(55cm) x 23″(60cm). 

4.) Keep the back and the sides of the autoclave approximately 10 cm away from the wall to allow ventilation and facilitate the device disconnection. 

5.) If placed in a cabinet, verify that the rear of the cabinet is open to allow ventilation. 

A caution symbol Caution! Insufficient space for ventilation may result in malfunction or damage due to overheating. 

6.) It is recommended that enough space be left around the autoclave to give a technician access for servicing the machine. 

7.) Check and verify that the ambient temperature range is 41oF (5oC)- 104oF(40oC), it is preferable not to exceed 86 oF (30oC). 

8.) Check and verify that the ambient relative humidity does not exceed 80% 

9.) The operational altitude shall not be over 13123 ft (4000 meters). 

10.) Ambient pressure shall not be lower than 8.8 psi (60.5 KPa) (if the altitude and temperature are kept in the manufacturer’s instructed ranges above, and no forced extreme negative pressure is applied near the autoclave, then ambient pressure of 8.8 psi (60.5 KPa) or higher is guaranteed, as it is a function of altitude and temperature). 

11.) Operate the autoclave only in the manner specified in the manual. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. 


Connections to Utility Supplies
  1. Check and verify that the power supply is a 1 phase, 120VAC ±10%, 60Hz, 15A supply. 
  2. Check and verify that the electrical net is protected by a current leakage safety relay. 

Storage
  • After the removal of the autoclave from the package, we recommend the following: 
  • Keep the device dry.
  • Keep the device away from sunlight and protect it from heat. 

Initial Operation of the Device

Note: Remove all packaging material before turning ON the device (see below).

This is the packing material removal graphic for the Tuttnauer T-Edge 11S.1.) Plug the cord into the power socket.

This is the power socket location on the Tuttnauer T-Edge 11S.

Warning! Mind the power socket. Keep its vicinity dry. Danger of electrocution.

2.)  Turn ON the semi-automatic ON/ OFF button switch located on the bottom left side of the back panel of the autoclave (see the rear view). 

This is the power switch location on the Tuttnauer T-Edge 11S.

3.) Turn on the ON/ OFF Switch mounted on the bottom left side of the front panel (see the front view).

This is the On/Off switch on the Tuttnauer T-Edge 11S.4.) When you turn on the autoclave, it will automatically warm up.

A caution symbol Warning! Be careful the surfaces may be hot!

5.) Fill the Mineral Free Water Reservoir with water meeting the quality specs.

6.) Set date and time.


Before Starting the Autoclave

Important! Reset the Atmospheric Pressure prior to operating the Autoclave for the first time. 

To reset the Atmospheric Pressure: 

1.) Open the Autoclave door and verify that the Chamber’s ambient temperature is below 113°F. 

2.) In the Quick Option screen, press the Settings icon, then browse to Maintenance/Reset atmospheric pressure.

3.) In the confirmation screen, press Reset

Before starting the autoclave we recommend that you check the following: 

1.) In the Quick Option screen, press the Info option. 

2.) Select the Version information.

3.) The Version information is displayed. For an example, see below: 

4.) Press on the to return to the Info screen.

5.) Select the Play tutorial option to watch the T-Edge 11 Instructional 



As always if you have any questions about this process or anything else please feel free to contact us and take advantage of our “FREE TECH SUPPORT.”

We also offer FREE VIRTUAL TECH SUPPORT to “See and Talk” with a “Real Time Live Technician” for any problems you may be in need of help with.

You can also use our “FREE MAINTENANCE PROGRAM”. Take the guesswork and worrying about what unit is due for maintenance and which maintenance cycle it is time for. We will keep track of all your autoclaves and let you know when it’s time for anything.


The Tuttnauer T-Edge 11S is available here on our site.

Parts for the Tuttnauer T-Edge line of autoclaves is also available on our site here.

SHOPPING CART

close