Enbio Pro Maintenance Guide

Proper installation, operation and maintenance are the most important things to know in terms of autoclave ownership. Improper installation can lead to the machineEnbio Pro top left front corner view not functioning properly. Improper operation can lead to complications with sterilization. And neglecting maintenance will ensure that your sterilizer will malfunction and cause long term damage.

But here at Sterilizer Autoclave Solutions, we make it our job to provide you with the proper knowledge to make sure everything goes smoothly for every phase of the process. In this article, we’d like to focus on the maintenance guide for one autoclave in particular. We’ll be going over the maintenance instructions for the Enbio Pro, with instructions provided by the manufacturer.

If you’d like to read the Installation Guide for the Enbio Pro, you can find it here. We also have an Operation Guide for the Enbio Pro on our site available to read here.

If you have any questions about the Enbio Pro autoclave or anything else, please give us a call at 704-966-1650 Option 3 for our Free Tech Support line.

 


Device Maintenance

In order to ensure the correct operation of the device, the user is obliged to perform the following maintenance actions.

Cleaning the Tray

Keeping the tray clean helps to ensure proper operation of the appliance. The tray and its technical condition are a good indicator of using the correct water. A stony, brown tray indicates the use of poor quality water. It is recommended to clean the inside of the tray on a weekly basis with a mild detergent that does not add chlorine and does not react with aluminum. After cleaning, the tray requires thorough rinsing with water. In order to clean the tray properly, it should be removed from the front of the device.

This is a guide for cleaning the Enbio Pro tray.

Cleaning the Process Chamber

Keeping the chamber clean helps to ensure proper operation of the device. It is recommended to clean the inside of the process chamber once a week with a mild detergent without the addition of chlorine. After cleaning, the chamber should be wiped dry with a soft cloth. Cleaning should be performed on a cold chamber.

Cleaning of External Surfaces

Cleaning the external parts of the device should be done with a soft cloth moistened with water and a mild detergent (no chlorine added and not reacting with plastics, varnish coatings, aluminum). Strong detergents should not be used. The use of mild detergents to maintain the device does not affect the risk of toxic components coming into contact with device components.

Cleaning the Gasket

It is recommended to clean the gasket each time after 100 processes. Use warm, clean water and microfiber to clean the gasket (microfiber with silver particles is allowed). Blunt and sharp cleaning tools are not allowed. Cleaning with chemicals is not allowed. Carry out cleaning on a cooled down device, after opening the drawer. Be careful not to bend the drawer. After cleaning, leave the device open until the gasket is dry. During this time, protect the device against damage.

Replacement of consumables

The sterilizer is equipped with a high performance sterilization system. A message on the screen informs the user when each component should be replaced. If the sterilizer is in regular operation, alternating replacement screens will appear after the welcome screen has been pressed. 

Cleaning the Water Container

In order to ensure the correct parameters of the water supplying the device, it is recommended to check the water tank at least quarterly. If contamination is found, the tank should be emptied, cleaned and refilled with new water.

In order to ensure an efficient sterilization process and the correct functioning of the device, it is recommended to replace consumables in a timely manner.

Replacing the Enbio Magic Filter

The filter should be replaced with a new one once every 6 months or sooner when the resin has completely discolored to an amber color. Failure to do so may disrupt the sterilization process and will void the sterilizer warranty. Dispose of the used filter according to local guidelines. The filter and all its parts are not recyclable.

This is the instruction guide to replace the Enbio Magic Filter

Consumables

The list below contains items subject to periodic replacement and items subject to natural wear. Spare parts must be ordered directly from the manufacturer. The use of other spare parts voids the warranty and does not guarantee the correct functioning of the device.

  • HEPA filter Enbio S / PRO: 1-8-27720A5
  • Enbio S front kit with gasket: 1-8-14672D2
  • Enbio S tray: 1-8-14738B2
  • Enbio PRO front kit with gasket: 1-8-12433A45
  • Tray set 3 pieces Enbio PRO 1-8-1097954A1
  • Wastewater set (plug + red hose 1.5 m): 1-8-42988A1
  • Feed water set (plug + blue hose 1.5 m): 1-8-42992A1
  • Enbio Magic Filter: 1-8-1116687A5

To ensure proper functioning of the Enbio S / Enbio PRO sterilizer, it is recommended that the wearing parts be replaced according to the schedule below. And periodic inspection of the individual components of the sterilizer in accordance with the following guidelines.

  • HEPA bacteriological filter – Every 1000 cycles or every 12 months
  • Connection / drain hose – If damage is observed or once a year
  • Plugs for water / condensate containers – If damage is observed
  • Front drawer with seal – Replacement at mandatory service after 1500 cycles/365 days
  • Enbio Magic Filter – Replace the filter once every 6 months or sooner when the resin has completely discolored to an amber color

Frequency of Inspections

  • Front gasket –  weekly or in the event of incorrect operation – performed by the user
  • Bacteriological filter – every week – performed by the user
  • Connection / drain hose – weekly or in the event of incorrect operation – performed by the user
  • Container stoppers – weekly – performed by the user
  • Enbio Magic Filter – weekly – performed by the user

 