What does the Midmark M9 Error Code E005 Mean?
Error E005: Pressure rises above 35.5 during the cycle, or M11 autoclave is overloaded.
Probable causes of Midmark M9 Error Code E005
- Pressure building in the chamber due to the bellows closing early
- Overloading the chamber
- A PC Board out of calibration (not an end user repair)
Solution: Stop the cycle, remove all instruments and start an empty cycle. If problem persists, contact Sterilizer Autoclave Solutions.
If you have any problems or are unsure, please use our FREE TECH SUPPORT to contact us.
View our Midmark M9 “Old Style” Error Code E005 Troubleshooting Video Below for More Help!