The W&H Assistina Twin Startup Guide

W&H Assistina Twin Startup Guide

As we expand our library of helpful content, we move on to a new subject. The W&H Assistina Twin is a new dual-chamber handpiece reprocessingW&H Assistina Twin handpiece processing device. machine. And a wonderful addition to your practice. We’ve written about the Assistina Twin here once before. But this article is going to be your W&H Assistina Twin Startup guide. The guide to setting up and preparing your Assistina Twin for the job ahead the way the manufacturer intended. If you have any questions, you can call us at 704-966-1650, option 3 for free tech support. 

 

Start-Up

Inserting/Removing a Cartridge

The cartridges must be at room temperature when they are inserted. 

A caution symbolThe cartridge spike in the medical device is sharp.
Be careful not to injure yourself on the projecting spike. 

Whenever you replace a cartridge:

  • Always clean the cartridge compartment
  • Perform the initial filling

Observe the manufacturer’s instructions on the use of disinfectant. 

Inserting/Removing a Filter

Always clean the filter compartment whenever you replace a filter. Observe the manufacturer’s instructions on the use of disinfectant. 

General

A caution symbolUse only non-contaminated, oil-free, dry and filtered compressed air according to ISO 7494-2. Replace the filter if contaminated. 

Switching the Assistina Twin On/Off

This is the W&H Assistina Twin switching power on/off graphic.

 

Initial Filling

Carry out the initial filling without a transmission instrument. It does not matter which chamber you fill first. 

Following the initial filling, clean both chambers with a damp cloth. 

Standby Mode

The medical device switches to standby mode: 

  • When the power supply unit is inserted into the socket; or
  • If it is not operated for 15 minutes. 

Test Run
  • Place a pre-disinfected transmission instrument onto the Quick Connect by W&H.
  • Start the Assistina Twin.

The transmission instrument will be slightly moist when you take it out. It then needs to be manually disinfected and sterilized. 

A caution symbol

In the event of operating malfunctions (e.g., vibrations, unusual noise, overheating, failure or leakage) stop the medical device immediately and contact an authorized W&H service partner.


As always if you have any questions about this process or anything else please feel free to contact us and take advantage of our “FREE TECH SUPPORT.”

We also offer FREE VIRTUAL TECH SUPPORT to “See and Talk” with a “Real Time Live Technician” for any problems you may be in need of help with.

You can also use our “FREE MAINTENANCE PROGRAM”. Take the guesswork and worrying about what unit is due for maintenance and which maintenance cycle it is time for. We will keep track of all your autoclaves and let you know when it’s time for anything.