This is the article on the SciCan Statim 2000 Cycle Fault 3.

SciCan Statim 2000 Cycle Fault 3

After certain periods of time, your autoclave may experience errors or serviceable issues. Through its use, your Statim 2000 can possibly display any number ofRefurbished Statim 2000 Cycle Fault codes that all pertain to different issues. Whether it’s repairs that you, the end-user can manage, or issues that may require a certified technician for further diagnosis. Like the Cycle Fault we’ll go over in this article, Cycle Fault 3. We’ll go over what the Cycle Fault means and any steps that can be taken as the end-user.

Note: Any step marked with “Field Service Tech. repair” is NOT meant to be serviced by the end-user. These steps are meant to be serviced by certified technicians. Any attempt to recreate technician repairs with improper tools or knowledge could result in damage to your autoclave.

What is Cycle Fault 3?

We have a video covering the steps required to service Cycle Fault 3 on the Statim 2000 here on our YouTube Channel.

For a visual approach, we have a video on the Statim 2000 Cycle Fault 3 on our YouTube channel that you can find here.

Cycle Fault 3 means: The cassette has failed to pressurize and achieve a temperature of 110°C within a time-out period.

  1. Check for visible steam leaks from the cassette. If the cassette is leaking repair as needed. Replace cassette seal, lid or trayEnd user repair
  2. If no leaks are visible disassemble the solenoid valve and check for debris or the plunger sticking in the plunger tube. Field Service Tech. repair
  3. Check for constant power to the solenoid. Measure AC voltage at J1-7 & J1-8, should be zero volts with cassette out of the unit and line voltage with cassette inserted in the unit. Field Service Tech. repair
  4. Verify that the check valve and pressure relief valves are not leaking. Field Service Tech. repair

As always if you have any questions about this process or anything else please feel free to contact us and take advantage of our “FREE TECH SUPPORT.”

We also offer FREE VIRTUAL TECH SUPPORT to “See and Talk” with a “Real Time Live Technician” for any problems you may be in need of help with.

You can also use our “FREE MAINTENANCE PROGRAM”. Take the guesswork and worrying about what unit is due for maintenance and which maintenance cycle it is time for. We will keep track of all your autoclaves and let you know when it’s time for anything.


The SciCan Statim 2000 can be found here on our website.

Parts for the SciCan Statim 2000 can be found here on our website.

Our SciCan YouTube playlist is available here for all things SciCan.

SHOPPING CART

close