Statim G4 Error Code Cycle Fault 16

What does Statim G4 Cycle Fault 16 Mean?

If you would like to talk to someone about this or any other questions you may have please call our Free Tech Support Line at 704-966-1650 Option 3.

 

The steam generator (boiler) temperature went above threshold value (171°C).

 

Probable Causes of Statim G4 Cycle Fault 16

1. Sticking Float Switch.
-To Repair This: Check water reservoir for water. If reservoir is empty, check to see if Float Switch is stuck in the raised position. Tap float with a long instrument, screw driver or ruler so float drops down. Refill reservoir with water. -This is an End-user repair

2. Bad Float Switch.
-To Repair This: If float continues to stick the Float Switch (part #01-110891S) in reservoir needs to be replaced. – This is a Field Service Tech repair

3. Malfunctioning Pump
-To Repair This: Check pump using the Pump Tester Bottle clean the pump filters if needed. – This is a Field Service Tech. repair

4. Boiler Power Issue
-To Repair This: Check for constant power to boiler and replace PCB if needed. – This is a Shop repair

View our Cycle Fault #16 Troubleshooting Video Below for More Help!