Statim G4 Error Code Cycle Fault 90

What does Statim G4 Cycle Fault 90 Mean?

If you would like to talk to someone about this or any other questions you may have please call our Free Tech Support Line at 704-966-1650 Option 3.

 

Corrupted or not initialized chamber calibration value.

 

Probable Causes of Statim G4 Cycle Fault 90

 

1. Corrupt NVRAM
-To Repair This: Go to Technician Menu/NVRAM Tools select restore NVRAM. -This is a Field Service Tech repair

2.Thermocouple Error
To Repair This: Calibrate Chamber thermocouple.– This is a Field Service Tech. repair

View our Cycle Fault #90 Troubleshooting Video Below for More Help!