Statim G4 Error Code Cycle Fault 12

What does Statim G4 Cycle Fault 12 Mean?

If you would like to talk to someone about this or any other questions you may have please call our Free Tech Support Line at 704-966-1650 Option 3.

 

This indicates a problem with the temperature measuring system.

 

Probable Causes of Statim G4 Cycle Fault 12

1. Bad Thermocouple
-To Repair This: Check thermocouples, they should read approximately 10 ohms at room temperature. -This is a Shop repair

View our Cycle Fault #12 Troubleshooting Video Below for More Help!

Get FREE Virtual Tech Support with Stanley Sterilizer!

Or Call Our Free Tech Support Line at 704-966-1650
We offer virtual technical help with our FREE Stanley Sterilizer Virtual Tech program.