Statim G4 Error Code Cycle Fault 8

What does Statim G4 Cycle Fault 8 Mean?

If you would like to talk to someone about this or any other questions you may have please call our Free Tech Support Line at 704-966-1650 Option 3.

 

The software has detected a Validation Thermocouple temperature 5°C less than the chamber temperature during the sterilizing phase of the cycle.

 

Probable Causes of Statim G4 Cycle Fault 8

1. Clogged Venturi
-To Repair This: Check for a clogged Venturi in the left rear of the cassette tray. Clean as needed. – This is an End-user repair

2. Dirty Solenoid
-To Repair This: Check the solenoid for debris and make sure the plunger is not sticking. – This is an End-user repair

3. Thermocouple Error
-To Repair This: Calibrate Validation thermocouple. – This is a Field Service Tech. repair