What does Statim G4 Cycle Fault 95 Mean?

If you would like to talk to someone about this or any other questions you may have please call our Free Tech Support Line at 704-966-1650 Option 3.

 

Communication error, the LCD controller PCB failed to receive messages from the Rev. 7 PCB.

 

Probable Causes of Statim G4 Cycle Fault 95

1. Voltage Issue.
-To Repair This: Check that 5V is available on Rev. 7 PCB and that Green light is lit on the front of the PCB. -This is a Field Service Tech repair

2. Bad Fuse.
-To Repair This: Check all fuses on Rev. 7 PCB. – This is a Field Service Tech repair

3. Cable Connection Issue.
-To Repair This: Check all cables for bad connection. -This is a Field Service Tech repair