What does Statim G4 Cycle Fault 4 Mean?

If you would like to talk to someone about this or any other questions you may have please call our Free Tech Support Line at 704-966-1650 Option 3.

 

The cassette has failed to pressurize and achieve a temperature of 110°C within a time- out period.

 

Probable Causes of Statim G4 Cycle Fault 4

1. Cassette Not Sealing Properly.
-To Repair This: Check for visible steam leaks from the cassette, if the cassette is leaking repair as needed. Replace cassette seal, lid or tray. -This is an End-user repair

2. Dirty Solenoid.
-To Repair This: Check the solenoid for debris and make sure the plunger is not sticking. – This is a Field Service Tech repair