Statim G4 Error Code Cycle Fault 7

What does Statim G4 Cycle Fault 7 Mean?

If you would like to talk to someone about this or any other questions you may have please call our Free Tech Support Line at 704-966-1650 Option 3.

 

The cassette temperature has dropped below a set point.

 

Probable Causes of Statim G4 Cycle Fault 7

 

If the cassette can be removed normally after venting.

1. A  Steam Leak
-To Repair This: Check for visible steam leaks from the cassette, if the cassette is leaking repair as needed. Replace cassette seal, lid or tray. -This is an End-user repair

2. Bad Solenoid
-To Repair This: Check the solenoid and make sure the plunger is not sticking. – This is an End-user repair

3. Bad Valve
-To Repair This: Verify that the Check Valve and Pressure Relief Valves are not leaking. – This is a Field Service Tech. repair

 

If the cassette is hard to remove after venting: (STATIM 2000 ONLY)

4. Bad Tubing
-To Repair This: Check for kinked or pinched exhaust tubing. – This is an End-user repair

5. Clogged Venturi
-To Repair This: Check for a clogged Venturi in the left rear of the cassette tray. Clean as needed. – This is an End-user repair

3. Bad Solenoid Valve
-To Repair This: Solenoid Valve is failing to open, Disassemble and check that the plunger slides smoothly in plunger tube. – This is a Field Service Tech. repair

View our Cycle Fault #7 Troubleshooting Video Below for More Help!