Statim G4 Error Code Cycle Fault 98

What does Statim G4 Cycle Fault 98 Mean?

If you would like to talk to someone about this or any other questions you may have please call our Free Tech Support Line at 704-966-1650 Option 3.

 

Square microprocessor failed to communicate with PCB.

 

Probable Causes of Statim G4 Cycle Fault 98

1. Power Surge.
-To Repair This: If power goes off and on during a cycle this cycle fault may appear. -This is an End-user or Field Service Tech repair

2. Microprocessor Failure.
-To Repair This: Check that pins on square microprocessor are not shorted. – This is a Field Service Tech repair

Get FREE Virtual Tech Support with Stanley Sterilizer!

Or Call Our Free Tech Support Line at 704-966-1650
We offer virtual technical help with our FREE Stanley Sterilizer Virtual Tech program.

Listed below are links to our articles on help for the other Statim G4 Error Codes.