What does Statim G4 Cycle Fault 96 Mean?

If you would like to talk to someone about this or any other questions you may have please call our Free Tech Support Line at 704-966-1650 Option 3.

 

Color LCD Controller program memory error, the LCD controller PCB fails to validate the integrity of the software at power up.

 

Probable Causes of Statim G4 Cycle Fault 96

1. Out of Date Software.
-To Repair This: Update Software. -This is a Field Service Tech repair

2. LCD Failure.
-To Repair This: Replace Color LCD Controller. – This is a Shop repair

Get FREE Virtual Tech Support with Stanley Sterilizer!

Or Call Our Free Tech Support Line at 704-966-1650