Statim G4 Error Code Cycle Fault 26

What does Statim G4 Cycle Fault 26 Mean?

If you would like to talk to someone about this or any other questions you may have please call our Free Tech Support Line at 704-966-1650 Option 3.

 

The sterilization phase has failed to start within 3 minutes of the cassette reaching the sterilization temperature.

NOTE: The unit has to fail on 3 consecutive cycles for CF26 to display. “CYCLE INTERRUPTED” will be displayed for the first 2 occurrences.

 

Probable Causes of Statim G4 Cycle Fault 26

1. Steam Leak
-To Repair This: Check for visible steam leaks from the cassette. If the cassette is leaking repair as needed, replace cassette seal, lid, or tray. -This is an End-user repair

2. Dirty Solenoid
-To Repair This: Check the solenoid for debris and make sure the plunger is not sticking. – This is an End-user repair

3. Thermocouple Error
-To Repair This: Calibrate Validation thermocouple. – This is a Field Service Tech. repair

View our Cycle Fault #26 Troubleshooting Video Below for More Help!