Statim G4 Error Code Cycle Fault 4

What does Statim G4 Cycle Fault 4 Mean?

If you would like to talk to someone about this or any other questions you may have please call our Free Tech Support Line at 704-966-1650 Option 3.

 

The cassette has failed to pressurize and achieve a temperature of 110°C within a time- out period.

 

Probable Causes of Statim G4 Cycle Fault 4

1. Cassette Not Sealing Properly.
-To Repair This: Check for visible steam leaks from the cassette, if the cassette is leaking repair as needed. Replace cassette seal, lid or tray. -This is an End-user repair

2. Dirty Solenoid.
-To Repair This: Check the solenoid for debris and make sure the plunger is not sticking. – This is a Field Service Tech repair

View our Cycle Fault #4 Troubleshooting Video Below for More Help!

Get FREE Virtual Tech Support with Stanley Sterilizer!

Or Call Our Free Tech Support Line at 704-966-1650
We offer virtual technical help with our FREE Stanley Sterilizer Virtual Tech program.

Listed below are links to our articles on help for the other Statim G4 Error Codes.

SHOPPING CART

close