What does Statim G4 Cycle Fault 4 Mean?
If you would like to talk to someone about this or any other questions you may have please call our Free Tech Support Line at 704-966-1650 Option 3.
The cassette has failed to pressurize and achieve a temperature of 110°C within a time- out period.
Probable Causes of Statim G4 Cycle Fault 4
1. Cassette Not Sealing Properly.
-To Repair This: Check for visible steam leaks from the cassette, if the cassette is leaking repair as needed. Replace cassette seal, lid or tray. -This is an End-user repair
2. Dirty Solenoid.
-To Repair This: Check the solenoid for debris and make sure the plunger is not sticking. – This is a Field Service Tech repair
View our Cycle Fault #4 Troubleshooting Video Below for More Help!
Get FREE Virtual Tech Support with Stanley Sterilizer!
Or Call Our Free Tech Support Line at 704-966-1650
Listed below are links to our articles on help for the other Statim G4 Error Codes.
Statim G4 Error Code Cycle Fault #1
Statim G4 Error Code Cycle Fault #3
Statim G4 Error Code Cycle Fault #6
Statim G4 Error Code Cycle Fault #7
Statim G4 Error Code Cycle Fault #8
Statim G4 Error Code Cycle Fault #10
Statim G4 Error Code Cycle Fault #11
Statim G4 Error Code Cycle Fault #12
Statim G4 Error Code Cycle Fault #15
Statim G4 Error Code Cycle Fault #16
Statim G4 Error Code Cycle Fault #19
Statim G4 Error Code Cycle Fault #25
Statim G4 Error Code Cycle Fault #26
Statim G4 Error Code Cycle Fault #27
Statim G4 Error Code Cycle Fault #90
Statim G4 Error Code Cycle Fault #94
Statim G4 Error Code Cycle Fault #95