Statim G4 Error Code Cycle Fault 6

What does Statim G4 Cycle Fault 6 Mean?

If you would like to talk to someone about this or any other questions you may have please call our Free Tech Support Line at 704-966-1650 Option 3.

 

The software has detected the Validation thermocouple temperature to be 5°C greater than the chamber during the sterilizing phase of a cycle.

 

Probable Causes of Statim G4 Cycle Fault 6

1. A  Steam Leak
-To Repair This: Check for visible steam leaks from the cassette, if the cassette is leaking repair as needed. Replace cassette seal, lid or tray. -This is an End-user repair

2. Bad Tubing
-To Repair This: Check for kinked exhaust tubing. – This is an End-user repair

3. Bad Solenoid
-To Repair This: Check for sticking solenoid. – This is a Field Service Tech. repair

4. Thermocouple Error
-To Repair This: Calibrate Validation thermocouple. – This is a Field Service Tech. repair

View our Cycle Fault #6 Troubleshooting Video Below for More Help!