What does Statim G4 Cycle Fault 6 Mean?
If you would like to talk to someone about this or any other questions you may have please call our Free Tech Support Line at 704-966-1650 Option 3.
The software has detected the Validation thermocouple temperature to be 5°C greater than the chamber during the sterilizing phase of a cycle.
Probable Causes of Statim G4 Cycle Fault 6
1. A Steam Leak
-To Repair This: Check for visible steam leaks from the cassette, if the cassette is leaking repair as needed. Replace cassette seal, lid or tray. -This is an End-user repair
2. Bad Tubing
-To Repair This: Check for kinked exhaust tubing. – This is an End-user repair
3. Bad Solenoid
-To Repair This: Check for sticking solenoid. – This is a Field Service Tech. repair
4. Thermocouple Error
-To Repair This: Calibrate Validation thermocouple. – This is a Field Service Tech. repair
View our Cycle Fault #6 Troubleshooting Video Below for More Help!
Get FREE Virtual Tech Support with Stanley Sterilizer!
Or Call Our Free Tech Support Line at 704-966-1650
Listed below are links to our articles on help for the other Statim G4 Error Codes.
Statim G4 Error Code Cycle Fault #1
Statim G4 Error Code Cycle Fault #3
Statim G4 Error Code Cycle Fault #4
Statim G4 Error Code Cycle Fault #7
Statim G4 Error Code Cycle Fault #8
Statim G4 Error Code Cycle Fault #10
Statim G4 Error Code Cycle Fault #11
Statim G4 Error Code Cycle Fault #12
Statim G4 Error Code Cycle Fault #15
Statim G4 Error Code Cycle Fault #16
Statim G4 Error Code Cycle Fault #19
Statim G4 Error Code Cycle Fault #25
Statim G4 Error Code Cycle Fault #26
Statim G4 Error Code Cycle Fault #27
Statim G4 Error Code Cycle Fault #90
Statim G4 Error Code Cycle Fault #94
Statim G4 Error Code Cycle Fault #95