Statim Parts
Showing 49–96 of 463 results
-
- Statim 2000 Parts (RPI)
Scican Statim 2000 AIR COMPRESSOR KIT RPI Part #SCK020 OEM Part #01-100560S
RPI Part #SCK020
OEM Part #01-100560SAIR COMPRESSOR KIT
- 28mA @ 115VAC; 50/60 HZ.
- Rated air flow: 0.177 cfm
- Rated pressure: 2.84 PSI
- Wire leads: #22 AWG x 22” long
- See Notes: below
Fits: Chassis
Model(s) This Part Fits: Statim 2000, Statim G4 2000Notes:
Service Tip #1: A leaky check valve can cause moisture or steam to seep inside a compressor thus causing the compressor to fail. To avoid this problem, replace the check valve when replacing the compressor and replace the check valve every 2 years.
Service Tip #2: – The OEM recommends replacing the Compressor Filters every 6 months or every 500 cycles (whichever comes first).
The RPI Advantage – The RPI tubing in the Check Valve Kit (RPI Part #SCK011) is translucent – making it easy to see a leaky check valve. For your convenience, RPI includes a check valve in the Air Compressor Kit (RPI Part #SCK020)!
Includes:
- Air Compressor
- Check Valve Kit (RPI Part #SCK011)
- Cable Ties (RPI Part #RPT083)
- Air Compressor Filter (RPI Part #SCF002) installed
- Mounting hardware
SKU: SCK020 - New, Statim 2000 Parts (RPI)
Scican Statim 2000 ALIGNMENT TOOL RPI Part #SCT052 OEM Part #01-106776S (Probe Alignment Jig)
New, Statim 2000 Parts (RPI)Scican Statim 2000 ALIGNMENT TOOL RPI Part #SCT052 OEM Part #01-106776S (Probe Alignment Jig)
RPI Part #SCT052
OEM Part #01-106776S (Probe Alignment Jig)ALIGNMENT TOOL
- RPI exclusive tool to easily align probe bracket steam ports to cassette in the field – It’s the RPI Advantage!
- Size: 1.50″ x .31″ x .25″
- Material: Aluminum
- Compact and easy to use
- Also included in: RPI Field Service Smart Kit® (RPI Part #SCK028)
- See Notes below
Fits: Cassette to Probe Bracket
Model(s) This Part Fits: Statim 2000, Statim 5000, Statim G4 2000, Statim G4 5000Notes: If using the RPI Alignment Tool (RPI Part #SCT052), the cassette seal must be removed before continuing with Alignment Tool installation.
SKU: SCT052-2000 -
- New, Statim 2000 Parts (RPI)
Scican Statim 2000 BIOLOGICAL FILTER RPI Part #SCF032 OEM OEM Part #01-102119S
New, Statim 2000 Parts (RPI)Scican Statim 2000 BIOLOGICAL FILTER RPI Part #SCF032 OEM OEM Part #01-102119S
RPI Part #SCF032
OEM Part #01-102119SBIOLOGICAL FILTER
- 3/16″ ID hose connector
- Size: 1.60″ W x 2.70″ lg. x .80″ thk.
- PM Service Tip: The OEM recommends replacing the Biological Filter every 6 months or every 500 cycles (whichever comes first).
- Also included in: Sterilizer PM Kits (RPI Part #SCK035) and (RPI Part #SCK036)
- See Notes below for Serial #’s this part fits
Fits: Air Compressor
Model(s) This Part Fits: Statim 2000, Statim 5000, Statim G4 2000, Statim G4 5000SKU: SCF032-2000 - New, Statim 2000 Parts (RPI)
Scican Statim 2000 BUBBLE LEVEL RPI Part #SCL012 OEM Part #01-100684S (Kit)
New, Statim 2000 Parts (RPI)Scican Statim 2000 BUBBLE LEVEL RPI Part #SCL012 OEM Part #01-100684S (Kit)
RPI Part #SCL012
OEM Part #01-100684S (Kit)BUBBLE LEVEL
- 1″ in dia. x 0.44″ tall with three #2 mounting holes
- Material: black plastic housing, glass window
- Includes Bubble Level Only
Fits: Housing Assembly
Model This Part Fits: Statim 2000SKU: SCL012-2000 -
- Statim 2000 Parts (OEM), New
Scican Statim 2000 Cassette Seal and Lube Kit OEM 01-100028S
DONT FORGET TO VIEW OUR INSTALLATION VIDEO BELOW!!!
FREE SHIPPING!!!
(1) Scican Statim 2000 Cassette Seal and Lube Kit OEM 01-100028S
SAVE MONEY – Purchase the FULL PM KIT
- PM Kit #1 Available Here Includes – Cassette Seal – Check Valve – Air Compressor Filter
- PM Kit #2 Available Here Includes – Cassette Seal – Check Valve – Air Compressor Filter – Biological Filter
- PM Kit #3 Available Here Includes – Cassette Seal – Air Compressor Filter
- PM Kit #4 Available Here Includes – Cassette Seal – Air Compressor Filter – Biological Filter
SKU: 01-100028S-2000 - New, Statim 2000 Parts (RPI)
Scican Statim 2000 CASSETTE SEAL LIQUID SOAP RPI Part #RPS287
RPI Part #RPS287
OEM Part # (OEM Part # Not Available)LIQUID SOAP
- 4 per package – 1/8 oz. foil pouches
- Use when maintaining or replacing Cassette Seal (RPI Part #SCS001 and RPI Part #SCS029)
- SDS Sheet available – see Notes below
Model(s) This Part Fits: Statim 2000, Statim 5000, Statim 900, Statim G4 2000, Statim G4 5000
Notes: Please see the Safety Data Sheet for this part. (Adobe PDF)
Also Included in:
- Cassette Seal (RPI Part #SCS001 and RPI Part #SCS029)
- Sterilizer PM Kit (RPI Part #SCK035 and RPI Part #SCK036)
- Field Service Smart Kit® (RPI Part #SCK028)
Also Avaiable:
- Sci-Dry™ (16 oz. Refill) (RPI Part #SCA046)
- Sci-Dry™ (2 oz.) (RPI Part #SCA054)
- Cleaning Kit (RPI Part #RPK791)
SKU: RPS287 - New, Statim 2000 Parts (RPI)
Scican Statim 2000 CASSETTE SEAL RPI Part #SCS001 OEM Part #01-100028S
RPI Part #SCS001
OEM Part #01-100028SCASSETTE SEAL
- Material: Black silicone
- Size: ~7-3/4″ x 11-3/4″
- Includes: Cassette Seal and (4 pouches) Liquid Soap (RPI Part #RPS287)
- See Notes below for “PM Service Tip”
Fits: Cassette lid
Model(s) This Part Fits: Statim 2000, Statim G4 2000Notes:
PM Service Tip: lube Cassette Seal (RPI Part #SCS001) with (1) pouch of Liquid Soap (RPI Part #RPS287) every 30 days. The OEM recommends replacing the Cassette Seal every 6 months or every 500 cycles (whichever comes first).
SKU: SCS001 - New, Statim 2000 Parts (RPI)
Scican Statim 2000 CHECK VALVE KIT RPI Part #SCK011 OEM OEM Part #01-100574S/01-1016275/9442740
New, Statim 2000 Parts (RPI)Scican Statim 2000 CHECK VALVE KIT RPI Part #SCK011 OEM OEM Part #01-100574S/01-1016275/9442740
RPI Part #SCK011
OEM Part #01-100574S/01-1016275/9442740CHECK VALVE KIT
- 1/8″ FPT
- Material: Brass
- High temperature seals
- 3/16” brass barb fitting (installed)
- See Notes below
Fits: Steam Generator/Boiler to the Air Compressor
Model(s) This Part Fits: Statim 2000, Statim G4 2000Notes:
Service Tip: A leaky check valve can cause moisture or steam to seep inside a compressor thus causing the compressor to fail. To avoid this problem, replace the check valve when replacing the compressor and replace the check valve every 2 years.
The RPI Advantage – The RPI tubing in the Check Valve Kit is translucent – making it easy to see a leaky check valve.
Includes:
- Check Valve
- (2x) Cable Ties (High Temp) (RPI Part #RPT480)
- Silicone Tubing (RPI Part #RPT373)
- Brass Fitting (installed)
- Threadlocker 545 (RPI Part #RPA369
SKU: SCK011-2000 -
- New, Statim 2000 Parts (RPI)
Scican Statim 2000 COIL (SOLENOID VALVE) RPI Part #SCC005 OEM Part #01-104303S
New, Statim 2000 Parts (RPI)Scican Statim 2000 COIL (SOLENOID VALVE) RPI Part #SCC005 OEM Part #01-104303S
RPI Part #SCC005
OEM Part #01-104303SCOIL (SOLENOID VALVE)
- 120/110VAC 50/60 Hz., 9W; lg.; 23-1/2″ lg. #22 AWG Teflon® wires
- Also available: Solenoid Valve (RPI Part #SCV064) and Solenoid Valve Repair Kit (RPI Part #SCK003)
Fits: RPI and the OEM’s Solenoid Valve
Model(s) This Part Fits: Statim 2000, Statim 5000, Statim G4 2000, Statim G4 5000SKU: SCC005-2000 - New, Statim 2000 Parts (RPI)
Scican Statim 2000 COMPRESSOR FILTER RPI Part #SCF002 OEM Part #01-100207S
New, Statim 2000 Parts (RPI)Scican Statim 2000 COMPRESSOR FILTER RPI Part #SCF002 OEM Part #01-100207S
RPI Part #SCF002
OEM Part #01-100207S / (OEM Part # Not Available)COMPRESSOR FILTER
- Size: 1.08″ ID x 2.08″ OD x .31″ Thk.
- Material: Polyethylene/Polyurethane Foam
- 40 micron rating
- Used with Air Compressor Kit (RPI Part #SCK020) and Pump (Scican Part #01-100560S or Pelton & Crane Part #3325086)
- Also included in: Air Compressor Kit (RPI Part #SCK020)
- PM Service Tip: The OEM recommends replacing the Compressor Filter every 6 months or every 500 cycles (whichever comes first).
Fits: Air Compressor
Model(s) This Part Fits: Delta 10″ (AF), Delta 8″ (AE), Delta Q10, Delta Q8, Delta XL10, Delta XL8, Statim 2000, Statim G4 2000, Statim G4 5000SKU: SCF002 - New, Statim 2000 Parts (OEM)
Scican Statim 2000 Compressor OEM 01-100560S
FREE SHIPPING!!!
(1) Scican Statim 2000 Compressor OEM 01-100560S
Notes:
Service Tip #1: A leaky check valve can cause moisture or steam to seep inside a compressor thus causing the compressor to fail. To avoid this problem, replace the check valve when replacing the compressor and replace the check valve every year.
Service Tip #2: – The OEM recommends replacing the Compressor Filters every 6 months or every 500 cycles (whichever comes first).
SKU: 01-100560S-2000 - New, Statim 2000 Parts (RPI)
Scican Statim 2000 CONDENSATION COIL RPI Part #SCC019
RPI Part #SCC019
OEM Part # (OEM Part # Not Available)CONDENSATION COIL
- 1/4” OD copper tubing
Fits: Condenser Waste Bottle Kit (RPI Part #SCK016) only
Model(s) This Part Fits: Statim 2000, Statim 5000, Statim 900, Statim G4 2000, Statim G4 5000SKU: SCC019-2000 - New, Statim 2000 Parts (RPI)
Scican Statim 2000 COPPER TUBING KIT RPI Part #SCK048 OEM Part #01-109128S
New, Statim 2000 Parts (RPI)Scican Statim 2000 COPPER TUBING KIT RPI Part #SCK048 OEM Part #01-109128S
RPI Part #SCK048
OEM Part #01-109128SCOPPER TUBING KIT
- Includes: (4 pcs) Compression Sleeve (RPI Part #RPF215), (4 pcs) Compression Nut (RPI Part #RPF219), (4) 3/16″ OD formed tubing
Fits: Probe Bracket to Boiler, Solenoid Valve and Alex Upgrade Kit
Model(s) This Part Fits: Statim 2000, Statim G4 2000SKU: SCK048 - Statim 2000 Parts (OEM)
SCICAN STATIM 2000 Copper Tubing OEM 01-109128S
FREE SHIPPING!!!
(1) NEW – SCICAN STATIM 2000 Copper Tubing OEM 01-109128S
COPPER TUBING KIT
Fits: Probe Bracket to Boiler, Solenoid Valve and Alex Upgrade Kit
Model(s) This Part Fits: Statim 2000, Statim G4 2000SKU: 01-109128S-2000 -
- Statim 2000 Parts (OEM)
SCICAN STATIM 2000 Cover Complete OEM 01-106121S
FREE SHIPPING!!!
(1) NEW – SCICAN STATIM 2000 Cover Complete OEM 01-106121S
Note: This is not a full unit, this includes the back cover and the facia to make a cover complete.
Model(s) This Part Fits: Statim 2000SKU: 01-106121S -
- New, Statim 2000 Parts (RPI)
Scican Statim 2000 DISPLAY ASSEMBLY RPI Part #SCA022 OEM Part #01-104382S/01-104381S (Display only)/01-109494S (Display only)
New, Statim 2000 Parts (RPI)Scican Statim 2000 DISPLAY ASSEMBLY RPI Part #SCA022 OEM Part #01-104382S/01-104381S (Display only)/01-109494S (Display only)
RPI Part #SCA022
OEM Part #01-104382S/01-104381S (Display only)/01-109494S (Display only)DISPLAY ASSEMBLY
- Display Assembly can be used in all versions of the STATIM 2000 and 5000. It’s the RPI Advantage!
- Includes: (1 pc) Mounting Pad (RPI Part #RPP819) and mounting hardware
- See Notes below
Fits: Facia Cover
Model(s) This Part Fits: Statim 2000, Statim 5000Notes:
Also Available:
- LCD Kit (RPI Part #SCK062)
- LCD Decal (RPI Part #SCD063)
- Keypad (RPI Part #SCK040)
- Keypad (RPI Part #SCK015)
- Keypad (RPI Part #SCK041)
SKU: SCA022-2000 - New, Statim 2000 Parts (RPI)
Scican Statim 2000 DOUBLE THERMAL FUSE ASSEMBLY RPI Part #SCF006 OEM Part #01-104461S/10-106462S
New, Statim 2000 Parts (RPI)Scican Statim 2000 DOUBLE THERMAL FUSE ASSEMBLY RPI Part #SCF006 OEM Part #01-104461S/10-106462S
RPI Part #SCF006
OEM Part #01-104461S/10-106462SDOUBLE THERMAL FUSE ASSEMBLY
- Temperature rating: Opens at 216°C
- Can be used on single fused steam generator/boilers
- Includes: Screw (#6-32×1/4) (RPI Part #ADS054) and (5pcs) Cable Tie (4″ White) (RPI Part #RPT083)
Fits: Steam Generator/Boiler
Model(s) This Part Fits: Statim 2000, Statim 5000, Statim 900, Statim G4 2000, Statim G4 5000SKU: SCF006-2000 - New, Statim 2000 Parts (RPI)
Scican Statim 2000 DRAIN KIT WITH PUMP FILTER RPI Part #SCK037 OEM Part #01-108152S (kit) & 01-106637S (Water Filter)
New, Statim 2000 Parts (RPI)Scican Statim 2000 DRAIN KIT WITH PUMP FILTER RPI Part #SCK037 OEM Part #01-108152S (kit) & 01-106637S (Water Filter)
RPI Part #SCK037
OEM Part #01-108152S (kit) & 01-106637S (Water Filter)DRAIN KIT WITH PUMP FILTER
- See Notes below for parts included in this kit
- See Notes below for “The RPI Advantage”
- PM Service Tip – The OEM recommends replacing the filter every 6 months or every 500 cycles (whichever comes first).
- Also available: Drain Tube Assembly (RPI Part #RPK432)
Fits: From the Reservoir to the Pump Filter to the Pump to the Drain Line
Model(s) This Part Fits: Statim 2000Notes: To upgrade the Scican Statim® 5000, order the Drain Tubing Kit (RPI Part #RPK432) and the Female Quick Disconnect Fitting (RPI Part #RPF428). The Statim 5000 comes equipped with the drain Line and Plug.
The RPI Advantage – The need for Pump Filter replacement is easily detected with RPI’s Clear Tubing and Filter. Draining is simplified with the addition of our Quick Disconnect System – Just Attach the tubing and you are ready to drain.
Includes:
- Cord Clip (RPI Part #ADC058)
- Barb Tee (RPI Part #RPB703)
- Cable Clamp (RPI Part #RPC701)
- Cable Holder (RPI Part #RPC702)
- Cable Tie (3 pieces) (RPI Part #RPT083)
- Drain Tubing Kit (RPI Part #RPK432)
- Female Quick Disconnect Fitting (RPI Part #RPF428)
- Hex Nut (RPI Part #RPH111)
- Silicone Tubing (RPI Part #RPT373)
- Water Filter (In-Line) (RPI Part #SCF034)
SKU: SCK037 - New, Statim 2000 Parts (RPI)
Scican Statim 2000 DRAIN TUBE ASSEMBLY KIT (SMALL) RPI Part #RPK432 OEM 3003683
New, Statim 2000 Parts (RPI)Scican Statim 2000 DRAIN TUBE ASSEMBLY KIT (SMALL) RPI Part #RPK432 OEM 3003683
RPI Part #RPK432
OEM Part #3003683DRAIN TUBE ASSEMBLY KIT (SMALL)
- Fitting: Acetal; Tubing: Clear Polyethylene (.25 O.D. x .035 Wall);~50″ lg.
- See Notes below for parts included in this kit
- Also included in: Drain Kit With Pump Filter (RPI Part #SCK037)
Fits: Front Panel/Female Quick Connect Fitting-White (RPI Part #RPF428)
Model(s) This Part Fits: Statim 2000Notes:
Includes:
- Quick Connect Male Fitting – White (RPI Part #RPF429)
- O-Ring (RPI Part #RPO306)
- (~50″ piece) Tubing (RPI Part #RPT276)
SKU: RPK432 -
- New, Statim 2000 Parts (RPI)
Scican Statim 2000 EXHAUST TUBING KIT RPI Part #SCK017 OEM Part #01-100204S
New, Statim 2000 Parts (RPI)Scican Statim 2000 EXHAUST TUBING KIT RPI Part #SCK017 OEM Part #01-100204S
RPI Part #SCK017
OEM Part #01-100204SEXHAUST TUBING KIT
- Includes: 8′ of Exhaust Tubing (RPI Part #RPT380) and (5 pcs) Cord Clip (RPI Part #ADC058)
- Tubing Size: 8 ft. lg; Nylon®; .250″ OD X .170″ ID
- Also included in: Condenser Waste Bottle Kit (RPI Part #SCK016)
- See Important Note below regarding the In-Line Quick Connect (RPI Part #RPF371)
Fits: Solenoid Valve Fitting to Waste Bottle Fitting
Model(s) This Part Fits: Statim 2000, Statim G4 2000Notes:
Also Available:
- In-Line Quick Disconnect (RPI Part #RPF371)
- Exhaust Tubing (RPI Part #RPT380)
- Push-In Elbow Fitting (RPI Part #RPF363)
Important Note The In-Line Quick Disconnect (RPI Part #RPF371) will not function properly with the OEM Teflon® Tubing. Only use with the supplied Nylon® Tubing (RPI Part #RPT380).
SKU: SCK017-2000