Steris Amsco Model V116 Steam Sterilizer Features Include:
- Florescent display for easy reading
- Ink on paper impact printer
- 16″ x 16″ x 26″ chamber
- Rack and 2 shelves
- Freestanding unit
- Vertical Door
- Pre-vacuum Steam Sterilizer
- Fast & efficient
- Emergency manual exhaust valve
- Non-lubricated, steam-activated door seal
- Help screens for programming/troubleshooting alarm conditions
Factory-Set Cycles and Cycle Values
Pre-Vacuum Sterilizer Cycles and Cycle Values
| Cycles | Sterilizer Temp. | Sterilizer Time | Dry Time | Recommended Load | Validation Standard |
|---|---|---|---|---|---|
| PREVAC | 270°F (132°C) | 4 MIN. | 5 MIN. | Single Fabric Pack | ST-8 |
| PREVAC | 270°F (132°C) | 4 MIN. | 20 MIN. | Double-wrapped instrument trays, max. weight of 17 lbs (7.7 kg) each. Fabric packs. | ST-8 |
| GRAVITY | 250°F (121°C) | 30 MIN | 15 MIN | Fabric packs. | ST-8 |
| LIQUID | 250°F (121°C) | 45 MIN | 0 MIN | ST-8 |
Steam Flush Pressure-Pulse Sterilizer Cycles and Cycle Values
| Cycles | Sterilizer Temp. | Sterilizer Time | Dry Time | Recommended Load | Validation Standard |
|---|---|---|---|---|---|
| WRAP/SFPP | 270°F (132°C) | 4 MIN. | 20 MIN. | Double-wrapped instrument trays, max. wt.: 17 lbs (7.7 kg) each. Non-porous Goods, only. | ST-8 |
| SFPP | 270°F (132°C) | 4 MIN. | 20 MIN. | Fabric Packs. | ST-8 |
| PREVAC | 270°F (132°C) | 4 MIN | 20 MIN | Double-wrapped instrument trays, max. wt.: 17 lbs (7.7 kg) each. Fabric Packs. | ST-8 |
| VACUUM | 250°F (121°C) | 30 MIN | 15 MIN | Fabric Packs | ST-8 |
| Test Cycles for All Units | Sterilizer Temp. | Sterilizer Time | Dry Time | Recommended Load | Validation Standard |
|---|---|---|---|---|---|
| Leak Test | 270°F (132°C) | N/A | N/A | N/A | ST-8 |
| SFPP | 270°F (132°C) | 3-1/2 MIN. | 1 MIN. | DART or Bowie-Dick Test Pack | ST-8 |
| PREVAC | 270°F (132°C) | 3 MIN | 1 MIN | N/A | N/A |