Notes:
Service Tip #1: A leaky check valve can cause moisture or steam to seep inside a compressor thus causing the compressor to fail. To avoid this problem, replace the check valve when replacing the compressor and replace the check valve every year.
Service Tip #2: – The OEM recommends replacing the Compressor Filters every 6 months or every 500 cycles (whichever comes first).