NEW (1) – Enbio PRO Gasket V3 ASM OEM 1211605A
Functions Include:
- Pressure & Steam Containment: It seals the sterilization drawer airtight to withstand high temperatures (up to 273°F/134°C) and high pressure during the Class B fractionated vacuum pulses.
- Error Code Prevention: A degraded or poorly seated door gasket allows steam or vacuum leaks. This directly triggers machine faults like Code 5 (Sterilization pressure too low) or vacuum failure alerts.
Routine Maintenance:
- According to official Enbio Maintenance Guidelines, the front gasket should be inspected weekly by the user for cracks, tears, or alignment issues and wiped down to ensure proper sealing