M11D Jumper Wire
Function: It’s primary purpose is to connect two high-limit safety thermostats together. This connection acts as a critical fail-safe to prevent the autoclave’s chamber from overheating if it accidentally runs dry.
Fits: Hi Limit Thermostat to Thermostat (Newer Style Units Only)
Model This Part Fits:
$16.00
FREE SHIPPING!!!
Midmark M11D RPI JUMPER WIRE RPI Part #MIJ328
Only logged in customers who have purchased this product may leave a review.

FREE SHIPPING!!!
Based on the EPA toxicity categorization of antimicrobial products that ranges from Category I (high toxicity) down to Category IV, OPTIM 33TB has the lowest toxicity category assigned of which precautionary warning label is not required.
This compares with Category II / III warning / caution statements for most leading brands of antimicrobial products.
FREE SHIPPING!!!
(1) Scican Statim 2000 Compressor OEM 01-100560S
Notes:
Service Tip #1: A leaky check valve can cause moisture or steam to seep inside a compressor thus causing the compressor to fail. To avoid this problem, replace the check valve when replacing the compressor and replace the check valve every year.
Service Tip #2: – The OEM recommends replacing the Compressor Filters every 6 months or every 500 cycles (whichever comes first).